Open official notice

Cardiac workstation

Official notice

16 Jul 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Cardiac workstation. The cardiac workstation can intermittently provide inaccurate data related to the patient's native heart rate. Healthcare providers were able to recognize that secondary review was needed because the patient’s clinical presentation and symptoms were not consistent with the elevated heart rate displayed by the Cardiac Workstation. Although the elevated heart rate may serve as an indicator to the clinician that the data is inaccurate, the issue also impacts the accuracy of the entire ECG report, including waveform data, measurements, and interpretive statements. This issue is related to the Wired Patient Interface Module PIM accessory used with the Cardiac Workstation. In rare cases, when the Cardiac Workstation is returned from Standby mode, the system may not set the ECG sampling rate in the Wired PIM correctly. This can cause the ECG waveform and interpretation to be inaccurate. Published 2026-07-16.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82332

Official notice

2026-07-16

The hazard

The cardiac workstation can intermittently provide inaccurate data related to the patient's native heart rate. Healthcare providers were able to recognize that secondary review was needed because the patient’s clinical presentation and symptoms were not consistent with the elevated heart rate displayed by the Cardiac Workstation. Although the elevated heart rate may serve as an indicator to the clinician that the data is inaccurate, the issue also impacts the accuracy of the entire ECG report, including waveform data, measurements, and interpretive statements. This issue is related to the Wired Patient Interface Module PIM accessory used with the Cardiac Workstation. In rare cases, when the Cardiac Workstation is returned from Standby mode, the system may not set the ECG sampling rate in the Wired PIM correctly. This can cause the ECG waveform and interpretation to be inaccurate.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

860439, 860441

Other

RA RA-82332

Model numbers

860439, 860441

Recall Numbers

82332

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - 860439

Product 2

Lot or serial number - Not applicable., Model or catalogue number - 860441

Hazard

The cardiac workstation can intermittently provide inaccurate data related to the patient's native heart rate. Healthcare providers were able to recognize that secondary review was needed because the patient’s clinical presentation and symptoms were not consistent with the elevated heart rate displayed by the Cardiac Workstation. Although the elevated heart rate may serve as an indicator to the clinician that the data is inaccurate, the issue also impacts the accuracy of the entire ECG report, including waveform data, measurements, and interpretive statements. This issue is related to the Wired Patient Interface Module PIM accessory used with the Cardiac Workstation. In rare cases, when the Cardiac Workstation is returned from Standby mode, the system may not set the ECG sampling rate in the Wired PIM correctly. This can cause the ECG waveform and interpretation to be inaccurate.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Philips Medizin Systeme Boblingen.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82332

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 16 Jul 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button