Open official notice

Ortho optix™ reader

Official notice

19 Apr 2022

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Ortho optix™ reader. Ortho Clinical Diagnostics received a complaint regarding an incorrectly displayed passing QC status for a test (see example in table below), without the site having processed the multiple QC kits configured for the test. Ortho confirmed that any test processed with multiple QC kits may encounter this issue. If one QC kit out of the multiple QC kits configured for the test has successfully passed, the software will automatically assign a passing QC status to the test even if the other QC kit configured for the test is not run or does not pass. Recall start date: Apr 11, 2022 Published 2022-04-19.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64093

Official notice

2022-04-19

The hazard

Ortho Clinical Diagnostics received a complaint regarding an incorrectly displayed passing QC status for a test (see example in table below), without the site having processed the multiple QC kits configured for the test. Ortho confirmed that any test processed with multiple QC kits may encounter this issue. If one QC kit out of the multiple QC kits configured for the test has successfully passed, the software will automatically assign a passing QC status to the test even if the other QC kit configured for the test is not run or does not pass. Recall start date: Apr 11, 2022

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

64093

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected Products - Ortho Optix™ Reader, Lot or Serial Number - Not applicable., Model or catalog number - 6842222

Hazard

Ortho Clinical Diagnostics received a complaint regarding an incorrectly displayed passing QC status for a test (see example in table below), without the site having processed the multiple QC kits configured for the test. Ortho confirmed that any test processed with multiple QC kits may encounter this issue. If one QC kit out of the multiple QC kits configured for the test has successfully passed, the software will automatically assign a passing QC status to the test even if the other QC kit configured for the test is not run or does not pass. Recall start date: Apr 11, 2022

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Ortho-clinical Diagnostics, Ortho Optix™ Reader.

Source & status

Tier

official

Category

Home & Consumer

Official classification

Type II

Recall details

Recall number

64093

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 19 Apr 2022. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button