Olympus hf-resection electrode

Official notice

13 Jul 2023

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Olympus hf-resection electrode. Olympus has initiated this recall after receiving complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. Recall start date: June 23, 2023. Published 2023-07-13.

What to do now

Contact the manufacturer if you require additional information.

Recall number

73949

Official notice

2023-07-13

The hazard

Olympus has initiated this recall after receiving complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. Recall start date: June 23, 2023.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

73949

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA22557C WA22657C

Product 2

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA22538C WA22651C

Product 3

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA22723S WA22621S WA22721S WA22623S

Product 4

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA22641S WA22666S WA22766S WA22741S

Product 5

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA47760S

Product 6

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA22521C WA22623C WA22621C WA22523C

Product 7

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA22632D

Product 8

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA22543S WA22305S WA22541S WA22566S WA22542S WA22544S WA22540S

Product 9

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA22655C WA22355A WA22355C

Product 10

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA47705S WA47737S WA47732S WA47707S WA47739S WA47742S WA47743S WA47706S

Product 11

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA22651S WA22751S WA22638S WA22738S

Product 12

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA22760S WA22657S

Product 13

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA47721S WA47723S

Product 14

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA47738S WA47751S

Product 15

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA47741S WA47766S

Product 16

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA47506S WA47540S WA47507S WA47560S WA47555S WA47505S WA47551S WA47566S

Product 17

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA22740S WA22755S WA22644S WA22655S WA22640S WA22744S

Product 18

Affected products - HF-Resection Electrode, Lot or serial number - All lots., Model or catalog number - WA47755S WA47740S WA47744S

Hazard

Olympus has initiated this recall after receiving complaints where the loop wire at the distal end of the HF resection electrode broke after getting in contact with metal objects, such as other endoscopic equipment, implants or stents. As a result, fragments can fall inside the patient and must be retrieved. Recall start date: June 23, 2023.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Olympus Winter Ibe, Olympus Hf-resection Electrode.

Source & status

Tier

official

Category

Health

Official classification

Type II

Recall details

Recall number

73949

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 13 Jul 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button