Image not available in RecallScope.
Pentax medical ultrasound endoscopes and water bottle assembly
Official notice
2 Oct 2026
Last checked
13 Nov 2026
An official Health Canada notice concerns Pentax medical ultrasound endoscopes and water bottle assembly. Pentax Medical has received an information related to a reported patient death in connection with a case in which air unintentionally entered the balloon attached to the distal end of an ultrasound endoscope during an endoscopic ultrasonography procedure.As a result of the investigation, a definitive causal relationship between the ingress of air into the balloon and the patient's death could not be established. However, it also cannot be completely ruled out. Therefore, to clearly address this potential risk, Pentax Medical has added a new warning to the IFU for the relevant ultrasound endoscopes and the water bottle assembly. This additional warning will be included in the IFU of all newly manufactured devices. For devices already placed on the market the users shall add this field safety notice to their documentation and inform their staff accordingly.As this issue may, under certain circumstances, potentially lead to a serious adverse event, Pentax Medical provides the following information to all customers using ultrasound endoscopes to ensure continued safe use of our products. Published 2026-10-02.
What to do now
Contact the manufacturer if you require additional information.
Recall number
82698
Official notice
2026-10-02
The hazard
The hazard
Pentax Medical has received an information related to a reported patient death in connection with a case in which air unintentionally entered the balloon attached to the distal end of an ultrasound endoscope during an endoscopic ultrasonography procedure.As a result of the investigation, a definitive causal relationship between the ingress of air into the balloon and the patient's death could not be established. However, it also cannot be completely ruled out. Therefore, to clearly address this potential risk, Pentax Medical has added a new warning to the IFU for the relevant ultrasound endoscopes and the water bottle assembly. This additional warning will be included in the IFU of all newly manufactured devices. For devices already placed on the market the users shall add this field safety notice to their documentation and inform their staff accordingly.As this issue may, under certain circumstances, potentially lead to a serious adverse event, Pentax Medical provides the following information to all customers using ultrasound endoscopes to ensure continued safe use of our products.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Skus
Other
Model numbers
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Product 3
Product 4
Product 5
Product 6
Product 7
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Pentax Medical.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.