Pentax medical ultrasound endoscopes and water bottle assembly

Official notice

2 Oct 2026

Last checked

13 Nov 2026

RecallScope risk estimate:HighHealth Canada

An official Health Canada notice concerns Pentax medical ultrasound endoscopes and water bottle assembly. Pentax Medical has received an information related to a reported patient death in connection with a case in which air unintentionally entered the balloon attached to the distal end of an ultrasound endoscope during an endoscopic ultrasonography procedure.As a result of the investigation, a definitive causal relationship between the ingress of air into the balloon and the patient's death could not be established. However, it also cannot be completely ruled out. Therefore, to clearly address this potential risk, Pentax Medical has added a new warning to the IFU for the relevant ultrasound endoscopes and the water bottle assembly. This additional warning will be included in the IFU of all newly manufactured devices. For devices already placed on the market the users shall add this field safety notice to their documentation and inform their staff accordingly.As this issue may, under certain circumstances, potentially lead to a serious adverse event, Pentax Medical provides the following information to all customers using ultrasound endoscopes to ensure continued safe use of our products. Published 2026-10-02.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82698

Official notice

2026-10-02

The hazard

Pentax Medical has received an information related to a reported patient death in connection with a case in which air unintentionally entered the balloon attached to the distal end of an ultrasound endoscope during an endoscopic ultrasonography procedure.As a result of the investigation, a definitive causal relationship between the ingress of air into the balloon and the patient's death could not be established. However, it also cannot be completely ruled out. Therefore, to clearly address this potential risk, Pentax Medical has added a new warning to the IFU for the relevant ultrasound endoscopes and the water bottle assembly. This additional warning will be included in the IFU of all newly manufactured devices. For devices already placed on the market the users shall add this field safety notice to their documentation and inform their staff accordingly.As this issue may, under certain circumstances, potentially lead to a serious adverse event, Pentax Medical provides the following information to all customers using ultrasound endoscopes to ensure continued safe use of our products.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

OS-H5, EG34-J10U, EG38-J10UT, EG-3270UK, EG36-J10UR, EG-3670URK, EG-3870UTK

Other

RA RA-82698

Model numbers

OS-H5, EG34-J10U, EG38-J10UT, EG-3270UK, EG36-J10UR, EG-3670URK, EG-3870UTK

Recall Numbers

82698

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - All lots., Model or catalogue number - OS-H5

Product 2

Lot or serial number - All lots., Model or catalogue number - EG34-J10U

Product 3

Lot or serial number - All lots., Model or catalogue number - EG38-J10UT

Product 4

Lot or serial number - All lots., Model or catalogue number - EG-3270UK

Product 5

Lot or serial number - All lots., Model or catalogue number - EG36-J10UR

Product 6

Lot or serial number - All lots., Model or catalogue number - EG-3670URK

Product 7

Lot or serial number - All lots., Model or catalogue number - EG-3870UTK

Hazard

Pentax Medical has received an information related to a reported patient death in connection with a case in which air unintentionally entered the balloon attached to the distal end of an ultrasound endoscope during an endoscopic ultrasonography procedure.As a result of the investigation, a definitive causal relationship between the ingress of air into the balloon and the patient's death could not be established. However, it also cannot be completely ruled out. Therefore, to clearly address this potential risk, Pentax Medical has added a new warning to the IFU for the relevant ultrasound endoscopes and the water bottle assembly. This additional warning will be included in the IFU of all newly manufactured devices. For devices already placed on the market the users shall add this field safety notice to their documentation and inform their staff accordingly.As this issue may, under certain circumstances, potentially lead to a serious adverse event, Pentax Medical provides the following information to all customers using ultrasound endoscopes to ensure continued safe use of our products.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Pentax Medical.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82698

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 2 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button