1688 4K pendulum camera head with integrated coupler

Official notice

8 Feb 2022

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns 1688 4K pendulum camera head with integrated coupler. Through investigation following a product compliant (November 11, 2021), Stryker has identified a software bug in the 1688 camera control unit (CCU) that can cause the image on the monitor to flip upside down (only) when used in conjunction with the 1688-310-130 Pendulum Camera Head. Recall start date: Jan 28, 2022 Published 2022-02-08.

What to do now

Contact the manufacturer if you require additional information.

Recall number

63908

Official notice

2022-02-08

The hazard

Through investigation following a product compliant (November 11, 2021), Stryker has identified a software bug in the 1688 camera control unit (CCU) that can cause the image on the monitor to flip upside down (only) when used in conjunction with the 1688-310-130 Pendulum Camera Head. Recall start date: Jan 28, 2022

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

63908

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected Products - 1688 4K PENDULUM CAMERA HEAD WITH INTEGRATED COUPLER, Lot or Serial Number - More than 10 numbers, contact manufacturer., Model or catalog number - 1688310130

Hazard

Through investigation following a product compliant (November 11, 2021), Stryker has identified a software bug in the 1688 camera control unit (CCU) that can cause the image on the monitor to flip upside down (only) when used in conjunction with the 1688-310-130 Pendulum Camera Head. Recall start date: Jan 28, 2022

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Stryker Endoscopy A Div Of Stryker Dba Stryker Endoscopy, 1688 4K Pendulum Camera Head With Integrated Coupler.

Source & status

Tier

official

Category

Health

Official classification

Type II

Recall details

Recall number

63908

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 8 Feb 2022. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button