Image not available in RecallScope.
1688 4K pendulum camera head with integrated coupler
Official notice
8 Feb 2022
An official Health Canada notice concerns 1688 4K pendulum camera head with integrated coupler. Through investigation following a product compliant (November 11, 2021), Stryker has identified a software bug in the 1688 camera control unit (CCU) that can cause the image on the monitor to flip upside down (only) when used in conjunction with the 1688-310-130 Pendulum Camera Head. Recall start date: Jan 28, 2022 Published 2022-02-08.
What to do now
Contact the manufacturer if you require additional information.
Recall number
63908
Official notice
2022-02-08
The hazard
The hazard
Through investigation following a product compliant (November 11, 2021), Stryker has identified a software bug in the 1688 camera control unit (CCU) that can cause the image on the monitor to flip upside down (only) when used in conjunction with the 1688-310-130 Pendulum Camera Head. Recall start date: Jan 28, 2022
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Stryker Endoscopy A Div Of Stryker Dba Stryker Endoscopy, 1688 4K Pendulum Camera Head With Integrated Coupler.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.