Molli 2 systems

Official notice

23 Jan 2026

RecallScope risk estimate:CriticalHealth Canada

An official Health Canada notice concerns Molli 2 systems. Stryker has updated the MOLLI 2 system Instructions for Use (IFU) following two medical device reports (MDRs) submitted to the FDA (June 5 2025). These reports involved patient injuries associated with MOLLI Marker dislodgement or dislocation due to magnetic attraction from magnetized surgical tools. The updated IFU now includes an additional warning that using magnetized surgical tools in close proximity to the marker during localization may lead to marker dislodgement. Published 2026-01-23.

What to do now

Contact the manufacturer if you require additional information.

Recall number

81524

Official notice

2026-01-23

The hazard

Stryker has updated the MOLLI 2 system Instructions for Use (IFU) following two medical device reports (MDRs) submitted to the FDA (June 5 2025). These reports involved patient injuries associated with MOLLI Marker dislodgement or dislocation due to magnetic attraction from magnetized surgical tools. The updated IFU now includes an additional warning that using magnetized surgical tools in close proximity to the marker during localization may lead to marker dislodgement.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

MP2015A, MT2001, MP2001, MS-00009, MM1000

Other

RA RA-81524

Model numbers

MP2015A, MT2001, MP2001, MS-00009, MM1000

Recall Numbers

81524

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Various, Model or catalogue number - MP2015A

Product 2

Lot or serial number - Various, Model or catalogue number - MT2001

Product 3

Lot or serial number - Various, Model or catalogue number - MP2001

Product 4

Lot or serial number - Various, Model or catalogue number - MS-00009

Product 5

Lot or serial number - Various, Model or catalogue number - MM1000

Hazard

Stryker has updated the MOLLI 2 system Instructions for Use (IFU) following two medical device reports (MDRs) submitted to the FDA (June 5 2025). These reports involved patient injuries associated with MOLLI Marker dislodgement or dislocation due to magnetic attraction from magnetized surgical tools. The updated IFU now includes an additional warning that using magnetized surgical tools in close proximity to the marker during localization may lead to marker dislodgement.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Stryker Endoscopy A Div Of Stryker Dba Stryker Endoscopy.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

81524

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 23 Jan 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button