Open official notice

UniCel DxI 800 Access Immunoassay Analyzer, Part Number 973100

Official notice

13 Aug 2025

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns UniCel DxI 800 Access Immunoassay Analyzer, Part Number 973100. Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed. Published 2025-08-13.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

2,203

Recall number

Z-2266-2025

Official notice

2025-08-13

The hazard

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 15099590265373

Event Ids

97162

Recall Numbers

Z-2266-2025

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

UniCel DxI 800 Access Immunoassay Analyzer, Part Number 973100

Units affected

2,203

Hazard

Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Beckman Coulter, Inc.

Phone

Not provided

Brand: Beckman Coulter. Contact: Beckman Coulter, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2266-2025

Event ID

97162

Units affected

2,203

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 13 Aug 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button