UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205
Official notice
13 Aug 2025
An official FDA notice concerns UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205. Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin. Published 2025-08-13.
What to do now
Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.
Units affected
56,696
Recall number
Z-2264-2025
Official notice
2025-08-13
The hazard
The hazard
Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Location 2
Location 3
Location 4
Location 5
Location 6
Location 7
Location 8
Location 9
Location 10
Affected area 1
Affected area 2
Affected area 3
Affected area 4
Affected area 5
Affected area 6
Affected area 7
Affected area 8
Affected area 9
Affected area 10
Affected area 11
Affected area 12
Affected area 13
Affected area 14
Affected area 15
Affected area 16
Affected area 17
Affected area 18
Affected area 19
Affected area 20
Affected area 21
Affected area 22
Affected area 23
Affected area 24
Affected area 25
Affected area 26
Affected area 27
Affected area 28
Affected area 29
Affected area 30
Affected area 31
Affected area 32
Affected area 33
Affected area 34
Affected area 35
Affected area 36
Affected area 37
Affected area 38
Affected area 39
Affected area 40
Affected area 41
Affected area 42
Affected area 43
Affected area 44
Affected area 45
Affected area 46
Company & contacts
Company & contacts
Brand: Beckman Coulter. Contact: Beckman Coulter Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.