Estrone ria, ref: dsl8700. Estrone ria is an ivd used for the quantitative measurement of estrone in human serum and plasma
Official notice
30 Jul 2025
An official FDA notice concerns Estrone ria, ref: dsl8700. Estrone ria is an ivd used for the quantitative measurement of estrone in human serum and plasma. A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits. Published 2025-07-30.
What to do now
Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.
Units affected
496
Recall number
Z-2152-2025
Official notice
2025-07-30
The hazard
The hazard
A specific lot of bovine serum albumin (BSA) used for manufacturing of coated tubes (CT) that are included in the Estrone RIA kit lots caused different affinity of manufactured coated tubes to patient samples, causing falsely increased assay results for the impacted Estrone RIA kits.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Lots
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Affected area 2
Affected area 3
Affected area 4
Affected area 5
Affected area 6
Affected area 7
Affected area 8
Affected area 9
Affected area 10
Affected area 11
Company & contacts
Company & contacts
Brand: Beckman Coulter. Contact: Beckman Coulter Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.