CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and

Official notice

5 Nov 2025

RecallScope risk estimate:CriticalFDA

An official FDA notice concerns CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and. Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment. Published 2025-11-05.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

108

Recall number

Z-0412-2026

Official notice

2025-11-05

The hazard

Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

97531

Recall Numbers

Z-0412-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek devices include Cassette E.

Units affected

108

Hazard

Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AT · Austria

Affected area 3

AU · Australia

Affected area 4

BE · Belgium

Affected area 5

BR · Brazil

Affected area 6

CA · Canada

Affected area 7

CH · Switzerland

Affected area 8

CN · China

Affected area 9

DE · Germany

Affected area 10

ES · Spain

Affected area 11

FR · France

Affected area 12

GB · United Kingdom

Affected area 13

GR · Greece

Affected area 14

HK · Hong Kong

Affected area 15

IL · Israel

Affected area 16

IT · Italy

Affected area 17

JP · Japan

Affected area 18

KR · South Korea

Affected area 19

MX · Mexico

Affected area 20

NL · Netherlands

Affected area 21

PL · Poland

Affected area 22

PT · Portugal

Affected area 23

ZA · South Africa

Company & contacts

Manufacturer

Beckman Coulter, Inc.

Phone

Not provided

Brand: Beckman Coulter. Contact: Beckman Coulter, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0412-2026

Event ID

97531

Units affected

108

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 5 Nov 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button