Open official notice

DxC 500 AU Module w ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i

Official notice

8 Oct 2025

RecallScope risk estimate:MediumFDA

An official FDA notice concerns DxC 500 AU Module w ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i. On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays. Published 2025-10-08.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

116

Recall number

Z-0032-2026

Official notice

2025-10-08

The hazard

On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

97390

Recall Numbers

Z-0032-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer

Units affected

116

Hazard

On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

BR · Brazil

Affected area 4

CA · Canada

Affected area 5

CH · Switzerland

Affected area 6

CL · Chile

Affected area 7

CZ · Czech Republic

Affected area 8

DE · Germany

Affected area 9

DO · Dominican Republic

Affected area 10

ES · Spain

Affected area 11

FR · France

Affected area 12

GB · United Kingdom

Affected area 13

HU · Hungary

Affected area 14

IN · India

Affected area 15

IT · Italy

Affected area 16

KR · South Korea

Affected area 17

MY · Malaysia

Affected area 18

NZ · New Zealand

Affected area 19

PL · Poland

Affected area 20

SA · Saudi Arabia

Affected area 21

SV · El Salvador

Affected area 22

TH · Thailand

Affected area 23

TR · Turkey

Company & contacts

Manufacturer

Beckman Coulter Inc.

Phone

Not provided

Brand: Beckman Coulter. Contact: Beckman Coulter Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0032-2026

Event ID

97390

Units affected

116

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 8 Oct 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button