Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer

Official notice

12 Nov 2025

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer. An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results. Published 2025-11-12.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

80

Recall number

Z-0428-2026

Official notice

2025-11-12

The hazard

An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

97730

Recall Numbers

Z-0428-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer

Units affected

80

Hazard

An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Beckman Coulter, Inc.

Phone

Not provided

Brand: Beckman Coulter. Contact: Beckman Coulter, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0428-2026

Event ID

97730

Units affected

80

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 12 Nov 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button