No image published for this recall

No image published for this recall

Puritan bennett 560 ventilator

Official notice

11 Jul 2024

Last checked

11 Jul 2024

Medium riskHealth Canada

An official Health Canada notice concerns Puritan bennett 560 ventilator. Standard toxicological test methods were employed to characterize the risk of exposure to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro, released from some components in the gas pathway of the device. Manufacturer determined that the device is safe for both intermittent and continuous use, for invasive and non-invasive ventilation for both pediatric and adult patients, over the entire duration of the labeled service life (10 years) and are in compliance with international published standards. For patients that require ventilation therapy beyond the full duration of the device's service life, a second unit may be required to allow the patient to continue receiving ventilation therapy beyond the initial 10 years. Additional test analyses suggest that cumulative use of over 14 years may potentially pose a harm to health. These conclusions are derived from chemical testing and risk assessment from literature, including animal laboratory testing. There are no conclusive human clinical data describing quantifiable risk. No customer complaints have been received reporting any of these potential harms. There are no anticipated harms associated with exposure to the identified chemical so long as the cumulative use of these devices does not exceed 14 years. Toxicological assessment based on small animal laboratory testing suggests that, with greater than 14 cumulative years of use of the device, exposure to the identified chemical has a maximum potential risk of 1 in 10,000 for cancer, birth defects/deformity, infertility, or abnormal fetal development. Continued, sequential, use of device is appropriate for patients who currently are using the devices for invasive and non-invasive ventilation therapy for up to and including 14 years of cumulative use. Any single device should only be used for the 10-year labelled service life, but if ventilatory support is required for a longer period using more than one device over time, the use of these devices for more than 14 years of cumulative duration is not advised. It is recommended that caregivers of patients approaching 14 years of cumulative use of device discuss changing their ventilatory support to an alternate ventilator with the patient's prescribing physician. Recall start date: June 27, 2024 Published 2024-07-11.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75808

Official notice

2024-07-11

The hazard

Standard toxicological test methods were employed to characterize the risk of exposure to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro, released from some components in the gas pathway of the device. Manufacturer determined that the device is safe for both intermittent and continuous use, for invasive and non-invasive ventilation for both pediatric and adult patients, over the entire duration of the labeled service life (10 years) and are in compliance with international published standards. For patients that require ventilation therapy beyond the full duration of the device's service life, a second unit may be required to allow the patient to continue receiving ventilation therapy beyond the initial 10 years. Additional test analyses suggest that cumulative use of over 14 years may potentially pose a harm to health. These conclusions are derived from chemical testing and risk assessment from literature, including animal laboratory testing. There are no conclusive human clinical data describing quantifiable risk. No customer complaints have been received reporting any of these potential harms. There are no anticipated harms associated with exposure to the identified chemical so long as the cumulative use of these devices does not exceed 14 years. Toxicological assessment based on small animal laboratory testing suggests that, with greater than 14 cumulative years of use of the device, exposure to the identified chemical has a maximum potential risk of 1 in 10,000 for cancer, birth defects/deformity, infertility, or abnormal fetal development. Continued, sequential, use of device is appropriate for patients who currently are using the devices for invasive and non-invasive ventilation therapy for up to and including 14 years of cumulative use. Any single device should only be used for the 10-year labelled service life, but if ventilatory support is required for a longer period using more than one device over time, the use of these devices for more than 14 years of cumulative duration is not advised. It is recommended that caregivers of patients approaching 14 years of cumulative use of device discuss changing their ventilatory support to an alternate ventilator with the patient's prescribing physician. Recall start date: June 27, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75808

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Puritan Bennett 560 Ventilator, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 4096600

Risk / hazard

Standard toxicological test methods were employed to characterize the risk of exposure to a volatile organic compound (VOC), 2-propanol, 1,3-dichloro, released from some components in the gas pathway of the device. Manufacturer determined that the device is safe for both intermittent and continuous use, for invasive and non-invasive ventilation for both pediatric and adult patients, over the entire duration of the labeled service life (10 years) and are in compliance with international published standards. For patients that require ventilation therapy beyond the full duration of the device's service life, a second unit may be required to allow the patient to continue receiving ventilation therapy beyond the initial 10 years. Additional test analyses suggest that cumulative use of over 14 years may potentially pose a harm to health. These conclusions are derived from chemical testing and risk assessment from literature, including animal laboratory testing. There are no conclusive human clinical data describing quantifiable risk. No customer complaints have been received reporting any of these potential harms. There are no anticipated harms associated with exposure to the identified chemical so long as the cumulative use of these devices does not exceed 14 years. Toxicological assessment based on small animal laboratory testing suggests that, with greater than 14 cumulative years of use of the device, exposure to the identified chemical has a maximum potential risk of 1 in 10,000 for cancer, birth defects/deformity, infertility, or abnormal fetal development. Continued, sequential, use of device is appropriate for patients who currently are using the devices for invasive and non-invasive ventilation therapy for up to and including 14 years of cumulative use. Any single device should only be used for the 10-year labelled service life, but if ventilatory support is required for a longer period using more than one device over time, the use of these devices for more than 14 years of cumulative duration is not advised. It is recommended that caregivers of patients approaching 14 years of cumulative use of device discuss changing their ventilatory support to an alternate ventilator with the patient's prescribing physician. Recall start date: June 27, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Covidien, Puritan Bennett 560 Ventilator.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

75808

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 11 Jul 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.