No image published for this recall

No image published for this recall

Mcgrath™ mac 2 laryngoscope blade

Official notice

18 Jan 2024

Last checked

18 Jan 2024

Medium riskHealth Canada

An official Health Canada notice concerns Mcgrath™ mac 2 laryngoscope blade. Medtronic received six (6) complaints of poor picture quality and foggy and/or blurred images on the McGRATH™ Mac 2 video laryngoscope screen during use of these blades. The reports indicated a delay to treatment while an alternate device was located and/or oral tissue injury due to multiple intubation attempts. Medtronic investigation of these customer reports identified a manufacturing issue which resulted in an ineffective application of the antifog agent on the blades. An ineffective application of the anti-fog coating can result in condensation accumulating on the blade during use, resulting in the image on the video laryngoscope screen being blurred or obscured. Recall start date: August 8, 2023 Published 2024-01-18.

What to do now

Contact the manufacturer if you require additional information.

Recall number

74924

Official notice

2024-01-18

The hazard

Medtronic received six (6) complaints of poor picture quality and foggy and/or blurred images on the McGRATH™ Mac 2 video laryngoscope screen during use of these blades. The reports indicated a delay to treatment while an alternate device was located and/or oral tissue injury due to multiple intubation attempts. Medtronic investigation of these customer reports identified a manufacturing issue which resulted in an ineffective application of the antifog agent on the blades. An ineffective application of the anti-fog coating can result in condensation accumulating on the blade during use, resulting in the image on the video laryngoscope screen being blurred or obscured. Recall start date: August 8, 2023

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

74924

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - McGRATH™ Mac 2 Laryngoscope Blade, Lot or serial number - 22102001 22082207 22083101, Model or catalog number - 350-017-000 350-084-000

Risk / hazard

Medtronic received six (6) complaints of poor picture quality and foggy and/or blurred images on the McGRATH™ Mac 2 video laryngoscope screen during use of these blades. The reports indicated a delay to treatment while an alternate device was located and/or oral tissue injury due to multiple intubation attempts. Medtronic investigation of these customer reports identified a manufacturing issue which resulted in an ineffective application of the antifog agent on the blades. An ineffective application of the anti-fog coating can result in condensation accumulating on the blade during use, resulting in the image on the video laryngoscope screen being blurred or obscured. Recall start date: August 8, 2023

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Covidien, Mcgrath™ Mac 2 Laryngoscope Blade.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

74924

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 18 Jan 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.