Nellcor™ bedside spo2 patient monitoring system
Official notice
25 Jun 2025
An official Health Canada notice concerns Nellcor™ bedside spo2 patient monitoring system. Medtronic is issuing a safety notification for its Nellcor™ Bedside SpO2 Patient Monitoring System. This notice follows receipt of multiple reports from customers stating that the device alarm was not heard or recognized, resulting in a delay to treatment or lack of response to low oxygen saturation, respiratory failure, or arrythmia. There have also been reports of death during some of these circumstances. Through investigation, Medtronic employees found that PM100N monitors were not set into homecare mode when in use outside of a hospital or other professional care setting. Homecare mode simplifies device operation and prevents lay users from making changes to patient mode and alarm setting. Remaining in standard mode allows for users to adjust the alarm limits and patient mode. Based on available information, the Nellcor™ bedside SpO2 patient monitors continues to be safe for use in patient care. While our investigation continues, this recall is being implemented to reiterate existing information available in the operator's and service manuals and the home use guide for both the PM100N and Libra 10005941 with possible additional activities initiated when the initial investigation has been completed. Recall start date: June 23, 2025 Published 2025-06-25.
What to do now
Contact the manufacturer if you require additional information.
Recall number
77648
Official notice
2025-06-25
The hazard
The hazard
Medtronic is issuing a safety notification for its Nellcor™ Bedside SpO2 Patient Monitoring System. This notice follows receipt of multiple reports from customers stating that the device alarm was not heard or recognized, resulting in a delay to treatment or lack of response to low oxygen saturation, respiratory failure, or arrythmia. There have also been reports of death during some of these circumstances. Through investigation, Medtronic employees found that PM100N monitors were not set into homecare mode when in use outside of a hospital or other professional care setting. Homecare mode simplifies device operation and prevents lay users from making changes to patient mode and alarm setting. Remaining in standard mode allows for users to adjust the alarm limits and patient mode. Based on available information, the Nellcor™ bedside SpO2 patient monitors continues to be safe for use in patient care. While our investigation continues, this recall is being implemented to reiterate existing information available in the operator's and service manuals and the home use guide for both the PM100N and Libra 10005941 with possible additional activities initiated when the initial investigation has been completed. Recall start date: June 23, 2025
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Covidien, Nellcor™ Bedside Spo2 Patient Monitoring System.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.