Image not available in RecallScope.
Merit basixcompak™, phd™ hemostasis valve, and basixsky™ inflation syringe
Official notice
14 Aug 2025
An official Health Canada notice concerns Merit basixcompak™, phd™ hemostasis valve, and basixsky™ inflation syringe. Merit Medical became aware of a manufacturing defect after an increased number of complaints. Investigation into the issue concluded that the lack of fluorinated silicone in the bleed back seal (the secondary valve) caused by the mix of incorrect silicone during production line influenced the performance of the PhD™ Hemostasis Valve. The lack of silicone was found to prevent the secondary valve from releasing and closing properly. The valve cap potentially would remain depressed causing leakage. Recall start date: July 30, 2025 Published 2025-08-14.
What to do now
Contact the manufacturer if you require additional information.
Recall number
77904
Official notice
2025-08-14
The hazard
The hazard
Merit Medical became aware of a manufacturing defect after an increased number of complaints. Investigation into the issue concluded that the lack of fluorinated silicone in the bleed back seal (the secondary valve) caused by the mix of incorrect silicone during production line influenced the performance of the PhD™ Hemostasis Valve. The lack of silicone was found to prevent the secondary valve from releasing and closing properly. The valve cap potentially would remain depressed causing leakage. Recall start date: July 30, 2025
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Product 3
Product 4
Product 5
Product 6
Product 7
Product 8
Product 9
Product 10
Product 11
Product 12
Product 13
Product 14
Product 15
Product 16
Product 17
Product 18
Product 19
Product 20
Product 21
Product 22
Product 23
Product 24
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Merit Medical, Merit Basixcompak™ Phd™ Hemostasis Valve And Basixsky™ Inflation Syringe.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.