Merit basixcompak™, phd™ hemostasis valve, and basixsky™ inflation syringe

Official notice

14 Aug 2025

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Merit basixcompak™, phd™ hemostasis valve, and basixsky™ inflation syringe. Merit Medical became aware of a manufacturing defect after an increased number of complaints. Investigation into the issue concluded that the lack of fluorinated silicone in the bleed back seal (the secondary valve) caused by the mix of incorrect silicone during production line influenced the performance of the PhD™ Hemostasis Valve. The lack of silicone was found to prevent the secondary valve from releasing and closing properly. The valve cap potentially would remain depressed causing leakage. Recall start date: July 30, 2025 Published 2025-08-14.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77904

Official notice

2025-08-14

The hazard

Merit Medical became aware of a manufacturing defect after an increased number of complaints. Investigation into the issue concluded that the lack of fluorinated silicone in the bleed back seal (the secondary valve) caused by the mix of incorrect silicone during production line influenced the performance of the PhD™ Hemostasis Valve. The lack of silicone was found to prevent the secondary valve from releasing and closing properly. The valve cap potentially would remain depressed causing leakage. Recall start date: July 30, 2025

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

77904

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - basixCOMPAK™ MAP802, Lot or serial number - I3161210, Model or catalog number - IN4802

Product 2

Affected products - basixCOMPAK™ MAP802, Lot or serial number - I3161211, Model or catalog number - IN4802

Product 3

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3146099, Model or catalog number - MAP801

Product 4

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3186946, Model or catalog number - MAP801

Product 5

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3219783, Model or catalog number - MAP802

Product 6

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3209935, Model or catalog number - MAP802

Product 7

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3215926, Model or catalog number - MAP802

Product 8

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3192937, Model or catalog number - MAP802

Product 9

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3215925, Model or catalog number - MAP802

Product 10

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3129726, Model or catalog number - MAP802

Product 11

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3129725, Model or catalog number - MAP802

Product 12

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3219784, Model or catalog number - MAP800

Product 13

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3206933, Model or catalog number - MAP800

Product 14

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3219792, Model or catalog number - MAP800

Product 15

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3186951, Model or catalog number - MAP800

Product 16

Affected products - PhD™ Hemostasis Valve, Lot or serial number - I3191334, Model or catalog number - MAP800

Product 17

Affected products - Custom Procedure Kit - Stroke Kit, Lot or serial number - T3163727, Model or catalog number - K12T-10798A

Product 18

Affected products - Custom Procedure Kit - Stroke Kit, Lot or serial number - T3217266, Model or catalog number - K12T-10798A

Product 19

Affected products - basixSKY™ Inflation Syringe MAP802, Lot or serial number - H3155575, Model or catalog number - SKY1802

Product 20

Affected products - basixSKY™ Inflation Syringe MAP802, Lot or serial number - H3141576, Model or catalog number - SKY1802

Product 21

Affected products - basixSKY™ Inflation Syringe MAP802, Lot or serial number - H3155576, Model or catalog number - SKY1802

Product 22

Affected products - basixSKY™ Inflation Syringe MAP802, Lot or serial number - H3155577, Model or catalog number - SKY1802

Product 23

Affected products - basixSKY™ Inflation Syringe MAP802, Lot or serial number - H3079946, Model or catalog number - SKY1802

Product 24

Affected products - basixSKY™ Inflation Syringe MAP802, Lot or serial number - H3141577, Model or catalog number - SKY1802

Hazard

Merit Medical became aware of a manufacturing defect after an increased number of complaints. Investigation into the issue concluded that the lack of fluorinated silicone in the bleed back seal (the secondary valve) caused by the mix of incorrect silicone during production line influenced the performance of the PhD™ Hemostasis Valve. The lack of silicone was found to prevent the secondary valve from releasing and closing properly. The valve cap potentially would remain depressed causing leakage. Recall start date: July 30, 2025

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Merit Medical, Merit Basixcompak™ Phd™ Hemostasis Valve And Basixsky™ Inflation Syringe.

Source & status

Tier

official

Category

Health

Official classification

Type II

Recall details

Recall number

77904

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Aug 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button