10Fore™ hemostasis valve

Official notice

30 Jul 2025

RecallScope risk estimate:HighHealth Canada

An official Health Canada notice concerns 10Fore™ hemostasis valve. Merit Medical became aware of a manufacturing quality concern during routine QA inspection of a 10Fore™ Hemostasis Valve work order in February 2025. Initial defect assessments determined the observed damage to the silicone quad ring was due to the manual nature of operator handling and subsequent placement/positioning of the quad ring into the corresponding y body. If the ring is not gently manipulated and/or not fully seated in its intended position during the pressing process it can become deformed or otherwise damaged. Recall start date: July 25, 2025 Published 2025-07-30.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77789

Official notice

2025-07-30

The hazard

Merit Medical became aware of a manufacturing quality concern during routine QA inspection of a 10Fore™ Hemostasis Valve work order in February 2025. Initial defect assessments determined the observed damage to the silicone quad ring was due to the manual nature of operator handling and subsequent placement/positioning of the quad ring into the corresponding y body. If the ring is not gently manipulated and/or not fully seated in its intended position during the pressing process it can become deformed or otherwise damaged. Recall start date: July 25, 2025

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

77789

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3098829, Model or catalog number - MAPTEN452

Product 2

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3111162, Model or catalog number - MAPTEN452

Product 3

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3098831, Model or catalog number - MAPTEN4

Product 4

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3111155, Model or catalog number - MAPTEN4

Product 5

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3152503, Model or catalog number - MAPTEN4

Product 6

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3155611, Model or catalog number - MAPTEN4

Product 7

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3186901, Model or catalog number - MAPTEN4

Product 8

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3076189, Model or catalog number - MAPTEN42

Product 9

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3111156, Model or catalog number - MAPTEN42

Product 10

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3155614, Model or catalog number - MAPTEN42

Product 11

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3173674, Model or catalog number - MAPTEN42

Product 12

Affected products - 10Fore™ Hemostasis Valve, Lot or serial number - H3152504, Model or catalog number - MAPTEN42

Hazard

Merit Medical became aware of a manufacturing quality concern during routine QA inspection of a 10Fore™ Hemostasis Valve work order in February 2025. Initial defect assessments determined the observed damage to the silicone quad ring was due to the manual nature of operator handling and subsequent placement/positioning of the quad ring into the corresponding y body. If the ring is not gently manipulated and/or not fully seated in its intended position during the pressing process it can become deformed or otherwise damaged. Recall start date: July 25, 2025

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Merit Medical, 10fore™ Hemostasis Valve.

Source & status

Tier

official

Category

Health

Official classification

Type I

Recall details

Recall number

77789

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 30 Jul 2025. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button