30 Atm monarch® inflation syringe
Official notice
9 Sept 2024
An official Health Canada notice concerns 30 Atm monarch® inflation syringe. On July 31, 2024, Merit internally identified an issue related to potential holes being found in the Tyvek® sterile barrier of certain MONARCH® inflation devices during the manufacture of lot H3003824. Visual inspection of the packaged units from lot H3003824 found a potential hole in 28 of 776 manufactured units. The 28 units underwent a bubble test where 23 of the 28 units failed, confirming a hole in the sterile barrier. Further investigation has determined that certain placements of the MONARCH® inflation device within a rigid formed blister tray can cause a hole in the Tyvek®. Should a MONARCH® inflation device's placement result in the unit's retainer cap being positioned above the rigid blister flange, the Tyvek® can contact a multivac's cross-cut station potentially leading to an abrasion or hole in the Tyvek®. Recall start date: August 28, 2024 Published 2024-09-09.
What to do now
Contact the manufacturer if you require additional information.
Recall number
76013
Official notice
2024-09-09
The hazard
The hazard
On July 31, 2024, Merit internally identified an issue related to potential holes being found in the Tyvek® sterile barrier of certain MONARCH® inflation devices during the manufacture of lot H3003824. Visual inspection of the packaged units from lot H3003824 found a potential hole in 28 of 776 manufactured units. The 28 units underwent a bubble test where 23 of the 28 units failed, confirming a hole in the sterile barrier. Further investigation has determined that certain placements of the MONARCH® inflation device within a rigid formed blister tray can cause a hole in the Tyvek®. Should a MONARCH® inflation device's placement result in the unit's retainer cap being positioned above the rigid blister flange, the Tyvek® can contact a multivac's cross-cut station potentially leading to an abrasion or hole in the Tyvek®. Recall start date: August 28, 2024
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Lots
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Merit Medical, 30 Atm Monarch® Inflation Syringe.
Source & status
Source & status
Recall details
Recall details
Recall number
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.