Open official notice

Freeo2 automated oxygen therapy device adult & pediatric - version hw 2.3

Official notice

8 Oct 2026

Last checked

13 Nov 2026

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Freeo2 automated oxygen therapy device adult & pediatric - version hw 2.3. Under specific conditions following a deep battery discharge, error 29515 may occur during startup of a FreeO2 HW 2.3 device using software package 1.4.2 and prevent its use. The update to software package 1.4.3 corrects this issue and restores the expected startup behaviour. Published 2026-10-08.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82749

Official notice

2026-10-08

The hazard

Under specific conditions following a deep battery discharge, error 29515 may occur during startup of a FreeO2 HW 2.3 device using software package 1.4.2 and prevent its use. The update to software package 1.4.3 corrects this issue and restores the expected startup behaviour.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

FO2-110-00

Other

RA RA-82749

Model numbers

FO2-110-00

Recall Numbers

82749

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Version HW 2.3, Model or catalogue number - FO2-110-00

Hazard

Under specific conditions following a deep battery discharge, error 29515 may occur during startup of a FreeO2 HW 2.3 device using software package 1.4.2 and prevent its use. The update to software package 1.4.3 corrects this issue and restores the expected startup behaviour.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada · Quebec

Company & contacts

Brand

Manufacturer

Not provided

Phone

Not provided

Brand: Oxynov.

Source & status

Tier

beta

Category

Health

Official classification

Type II

Recall details

Recall number

82749

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 8 Oct 2026. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.