Open official notice

VITEK 2 Gram Negative test kits containing colistin (cs01n)

Official notice

2 Aug 2017

RecallScope risk estimate:MediumFDA

An official FDA notice concerns VITEK 2 Gram Negative test kits containing colistin (cs01n). The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK¿ 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI). Published 2017-08-02.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

1,520,000

Recall number

Z-2770-2017

Official notice

2017-08-02

The hazard

The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK¿ 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI).

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

77373

Recall Numbers

Z-2770-2017

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VITEK 2 Gram Negative test kits containing colistin (cs01n)

Units affected

1,520,000

Hazard

The investigation demonstrated a high rate of very major errors (resistant isolates calling susceptible) with the VITEK¿ 2 AST-GN colistin (cs01n) compared to agar dilution (the reference method used for cs01n development) and compared to broth-microdilution (recommended method by EUCAST/CLSI).

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · NC

Affected area 1

US · United States

Affected area 2

AR · Argentina

Affected area 3

AT · Austria

Affected area 4

BR · Brazil

Affected area 5

CA · Canada

Affected area 6

CL · Chile

Affected area 7

CN · China

Affected area 8

CO · Colombia

Affected area 9

CR · Costa Rica

Affected area 10

DE · Germany

Affected area 11

DO · Dominican Republic

Affected area 12

ES · Spain

Affected area 13

FR · France

Affected area 14

GB · United Kingdom

Affected area 15

GT · Guatemala

Affected area 16

HK · Hong Kong

Affected area 17

IL · Israel

Affected area 18

IN · India

Affected area 19

IT · Italy

Affected area 20

JP · Japan

Affected area 21

KR · South Korea

Affected area 22

MX · Mexico

Affected area 23

MY · Malaysia

Affected area 24

NL · Netherlands

Affected area 25

PE · Peru

Affected area 26

PL · Poland

Affected area 27

SG · Singapore

Affected area 28

TH · Thailand

Affected area 29

TR · Turkey

Affected area 30

UY · Uruguay

Affected area 31

VE · Venezuela

Affected area 32

VN · Vietnam

Affected area 33

ZA · South Africa

Company & contacts

Manufacturer

Biomerieux Inc

Phone

Not provided

Brand: Biomerieux. Contact: Biomerieux Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2770-2017

Event ID

77373

Units affected

1,520,000

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 2 Aug 2017. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button