Open official notice

EMAG system Model Catalog Number: 418591 418591U Product Description: Nucleic acid sample preparation instrument IVD, automated

Official notice

4 Nov 2026

Last checked

10 Nov 2026

RecallScope risk estimate:MediumFDA

An official FDA notice concerns EMAG system Model Catalog Number: 418591 418591U Product Description: Nucleic acid sample preparation instrument IVD, automated. Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2 Published 2026-11-04.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

33

Recall number

Z-0301-2027

Official notice

2026-11-04

The hazard

Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

99931

Recall Numbers

Z-0301-2027

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated

Units affected

33

Hazard

Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AT · Austria

Affected area 3

BE · Belgium

Affected area 4

CA · Canada

Affected area 5

DE · Germany

Affected area 6

DK · Denmark

Affected area 7

FR · France

Affected area 8

GB · United Kingdom

Affected area 9

NL · Netherlands

Affected area 10

NO · Norway

Affected area 11

SE · Sweden

Company & contacts

Manufacturer

BioMerieux SA

Phone

Not provided

Brand: Biomerieux. Contact: BioMerieux SA.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0301-2027

Event ID

99931

Units affected

33

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 4 Nov 2026. The next one won’t wait for you to search.