Open official notice

BioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, V2S 6.01 RP5800-WES7 Media. Part

Official notice

14 Aug 2013

RecallScope risk estimate:MediumFDA

An official FDA notice concerns BioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, V2S 6.01 RP5800-WES7 Media. Part. A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bid-directional Computer interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin and Ceftaroline. For the referenced antibiotics, the firm code in the reference table Published 2013-08-14.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

12,384

Recall number

Z-1907-2013

Official notice

2013-08-14

The hazard

A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bid-directional Computer interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin and Ceftaroline. For the referenced antibiotics, the firm code in the reference table

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

65470

Recall Numbers

Z-1907-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

bioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, V2S 6.01 RP5800-WES7 Media. Part number: 413861. The product is used with the VITEK 2 automated system which consists of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.

Units affected

12,384

Hazard

A product issue has been identified for customers using software version 6.01 and are interfaced with a Laboratory Information System (LIS). During the update of software from version 5.xx to 6.01, the Bid-directional Computer interface (BCI) antibiotic codes are not updated properly for Cefovecin, Prulifloxacin and Ceftaroline. For the referenced antibiotics, the firm code in the reference table

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

AU · Australia

Affected area 6

BG · Bulgaria

Affected area 7

BR · Brazil

Affected area 8

CA · Canada

Affected area 9

CH · Switzerland

Affected area 10

CL · Chile

Affected area 11

CN · China

Affected area 12

CR · Costa Rica

Affected area 13

CZ · Czech Republic

Affected area 14

DE · Germany

Affected area 15

DO · Dominican Republic

Affected area 16

EG · Egypt

Affected area 17

ES · Spain

Affected area 18

FR · France

Affected area 19

GB · United Kingdom

Affected area 20

HK · Hong Kong

Affected area 21

HR · Croatia

Affected area 22

HU · Hungary

Affected area 23

IL · Israel

Affected area 24

IN · India

Affected area 25

IT · Italy

Affected area 26

JP · Japan

Affected area 27

KR · South Korea

Affected area 28

KW · Kuwait

Affected area 29

LT · Lithuania

Affected area 30

MT · Malta

Affected area 31

MX · Mexico

Affected area 32

MY · Malaysia

Affected area 33

NL · Netherlands

Affected area 34

PE · Peru

Affected area 35

PK · Pakistan

Affected area 36

PL · Poland

Affected area 37

PT · Portugal

Affected area 38

RO · Romania

Affected area 39

RS · Serbia

Affected area 40

RU · Russia

Affected area 41

SE · Sweden

Affected area 42

SG · Singapore

Affected area 43

SI · Slovenia

Affected area 44

TH · Thailand

Affected area 45

UY · Uruguay

Affected area 46

VN · Vietnam

Affected area 47

ZA · South Africa

Company & contacts

Manufacturer

Biomerieux Inc

Phone

Not provided

Brand: Biomerieux. Contact: Biomerieux Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1907-2013

Event ID

65470

Units affected

12,384

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 14 Aug 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button