MYLA comprises AST Filters in conjunction with VITEK MS; middleware which connects microbiology laboratory instruments for ID AST and blood

Official notice

8 Mar 2023

RecallScope risk estimate:MediumFDA

An official FDA notice concerns MYLA comprises AST Filters in conjunction with VITEK MS; middleware which connects microbiology laboratory instruments for ID AST and blood. For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician. Published 2023-03-08.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

19

Recall number

Z-1201-2023

Official notice

2023-03-08

The hazard

For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

91617

Recall Numbers

Z-1201-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

MYLA comprises AST Filters in conjunction with VITEK MS; middleware which connects microbiology laboratory instruments for ID/AST and blood culture activities and the laboratory information system (LIS).

Units affected

19

Hazard

For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Biomerieux Inc

Phone

Not provided

Brand: Biomerieux. Contact: Biomerieux Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1201-2023

Event ID

91617

Units affected

19

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 8 Mar 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button