V tome x L Series non-destructive testing cabinet x-ray systems

Official notice

25 Nov 2020

RecallScope risk estimate:MediumFDA

An official FDA notice concerns V tome x L Series non-destructive testing cabinet x-ray systems. It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm up" mode was not completed after initializing the system when first turned on. Published 2020-11-25.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Recall number

Z-0442-2021

Official notice

2020-11-25

The hazard

It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm up" mode was not completed after initializing the system when first turned on.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

86620

Recall Numbers

Z-0442-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

v|tome|x L Series non-destructive testing cabinet x-ray systems

Hazard

It was discovered during testing of the v/tome/x L450 system s/n PA1790, that the pre-warning signals does not provide "Audible and visible warning signals within the cabinet" if the "warm up" mode was not completed after initializing the system when first turned on.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CA · Canada

Company & contacts

Manufacturer

Ge Inspection Technologies Llc

Phone

Not provided

Brand: GE Inspection. Contact: Ge Inspection Technologies Llc.

Source & status

Source

Tier

official

Category

ELECTRONICS

Official classification

Class II

Recall details

Recall number

Z-0442-2021

Event id

86620

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 25 Nov 2020. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button