Open official notice

RT MonteCarlo 3D v3

Official notice

23 Sept 2026

Last checked

29 Sept 2026

RecallScope risk estimate:MediumFDA

An official FDA notice concerns RT MonteCarlo 3D v3. When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth. Published 2026-09-23.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

2

Recall number

Z-3174-2026

Official notice

2026-09-23

The hazard

When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Skus

S070031

Event Ids

99720

Model Numbers

S070031

Recall Numbers

Z-3174-2026

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VERIQA; Product Name: RT MonteCarlo 3D v3; Model/Catalog Number: S070031.520; Software Version: 3.1.18

Units affected

2

Hazard

When a bolus is included in a treatment plan and sent for secondary dose calculation, the software accounts for it but does not display this clearly, which may lead users to incorrectly omit the bolus during treatment and deliver the prescribed dose to the wrong tissue depth.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AT · Austria

Affected area 4

BE · Belgium

Affected area 5

BR · Brazil

Affected area 6

CH · Switzerland

Affected area 7

CL · Chile

Affected area 8

CZ · Czech Republic

Affected area 9

DE · Germany

Affected area 10

ES · Spain

Affected area 11

FR · France

Affected area 12

HK · Hong Kong

Affected area 13

MX · Mexico

Affected area 14

NL · Netherlands

Affected area 15

NO · Norway

Affected area 16

PL · Poland

Affected area 17

PT · Portugal

Affected area 18

RO · Romania

Affected area 19

RS · Serbia

Affected area 20

SA · Saudi Arabia

Affected area 21

TH · Thailand

Affected area 22

ZA · South Africa

Company & contacts

Manufacturer

PTW FREIBURG

Phone

Not provided

Brand: Ptw Freiburg. Contact: PTW FREIBURG.

Source & status

Source

Tier

official

Category

ELECTRONICS

Official classification

Class II

Recall details

Recall number

Z-3174-2026

Event id

99720

Units affected

2

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 23 Sept 2026. The next one won't wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button