Ultima OPCAB System, Sterile, Rx Only, Product Usage: Intended for use during performance of minimally invasive cardiac surgery through a

Official notice

9 Oct 2013

RecallScope risk estimate:MediumFDA

Ultima OPCAB System, Sterile, Rx Only, Product Usage: Intended for use during performance of minimally invasive cardiac surgery through a sternotomy incision approach. The ULTIMA Access RailPlatform in combination with the Activator II Drive Mechanism is used to spread the sternum, providing access and direct visualization to the thoracic cavity. The AccessRail Platform also allows for the organization of pericardial sutures. The ULTIMA Stablizer isolates and provides local immobilization of the target vessel on the beating heart.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

20

Recall number

Z-0001-2014

Official notice

2013-10-09

The hazard

MAQUET has identified a potential issue with the ULTIMA OPCAB (part numbers OM-2003S & OM-2001D). Labeling for the 2 blade lots had been incorrectly applied. The standard blades from Lot #25052176 were incorrectly labeled as deep blades, and the 20 deep blades from Lot #25052200 were incorrectly labeled as standard blades.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

25052176, 25052200

Event Ids

66365

Recall Numbers

Z-0001-2014

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Ultima OPCAB System, Sterile, Rx Only, Product Usage: Intended for use during performance of minimally invasive cardiac surgery through a sternotomy incision approach. The ULTIMA Access RailPlatform in combination with the Activator II Drive Mechanism is used to spread the sternum, providing access and direct visualization to the thoracic cavity. The AccessRail Platform also allows for the organization of pericardial sutures. The ULTIMA Stablizer isolates and provides local immobilization of the target vessel on the beating heart.

Units affected

20

Hazard

MAQUET has identified a potential issue with the ULTIMA OPCAB (part numbers OM-2003S & OM-2001D). Labeling for the 2 blade lots had been incorrectly applied. The standard blades from Lot #25052176 were incorrectly labeled as deep blades, and the 20 deep blades from Lot #25052200 were incorrectly labeled as standard blades.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Manufacturer

Maquet Cardiovascular, LLC

Phone

Not provided

Brand: Maquet Cardiovascular. Contact: Maquet Cardiovascular, LLC.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0001-2014

Event ID

66365

Units affected

20

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 9 Oct 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button