Open official notice

The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope. The

Official notice

17 Jul 2024

RecallScope risk estimate:MediumFDA

The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

600

Recall number

Z-2337-2024

Official notice

2024-07-17

The hazard

There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Other

UDI-DI 00607567700345

Event Ids

94802

Recall Numbers

Z-2337-2024

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

The VASOVIEW HEMOPRO Endoscopic Vessel Harvesting Systems is designed for use in conjunction with the 7 mm Extended Length Endoscope. The Harvesting Cannula has four lumens which house the Endoscope, C-Ring, distal lens washer tube, and harvesting tool for cutting and cauterizing vessel branches. The harvesting tool has two curved jaws. One of the jaws contains the heating elements for branch cutting and cauterization and spot cautery. An area near the tip of the convex side of the jaw is used for spot cautery. The activation toggle is used to control the jaws to activate the heating elements. Positioning of the device, cutting, and cauterization are performed under endoscopic visualization.

Units affected

600

Hazard

There were 27 complaints between March 22, 2024, and April 30, 2024, related to VH-3500 and VH-4000, reporting that the C-Ring wire was straightened (due to insufficient bend radius) and that the distance between C-Ring and Harvesting Tool was closer than normal. No adverse events have been reported to date in direct association with this issue.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · DE

Affected area 1

US · United States

Company & contacts

Manufacturer

Maquet Cardiovascular, LLC

Phone

Not provided

Brand: Maquet Cardiovascular. Contact: Maquet Cardiovascular, LLC.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-2337-2024

Event ID

94802

Units affected

600

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 17 Jul 2024. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button