MAQUET CARDIOHELP Support System: CARDIOHELP-I Software (up to version 3.3.0.0) Product Usage: The Cardiohelp System is a blood oxygenation

Official notice

24 Apr 2013

RecallScope risk estimate:MediumFDA

MAQUET CARDIOHELP Support System: CARDIOHELP-I Software (up to version 3.3.0.0) Product Usage: The Cardiohelp System is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support during procedures requiring cardiopulmonary bypass (for periods up to six hours). It is also intended to provide circulatory and/or pulmonary support during procedures not requiring cardiopulmonary bypass (for periods up to six hours).

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

154

Recall number

Z-1149-2013

Official notice

2013-04-24

The hazard

It has come to the attention of MAQUET that there have been episodes of brief unexpected shutdown, followed by an automatic device restart of the human machine interface (touch screen) of the CARDIOHELP device. The human machine interface (HMI) is the central display of the CARDIOHELP device, where measured values, settings and alarms are displayed and adjusted/confirmed, with the exception of bl

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

64641

Recall Numbers

Z-1149-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

MAQUET CARDIOHELP Support System: CARDIOHELP-I Software (up to version 3.3.0.0) Product Usage: The Cardiohelp System is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support during procedures requiring cardiopulmonary bypass (for periods up to six hours). It is also intended to provide circulatory and/or pulmonary support during procedures not requiring cardiopulmonary bypass (for periods up to six hours).

Units affected

154

Hazard

It has come to the attention of MAQUET that there have been episodes of brief unexpected shutdown, followed by an automatic device restart of the human machine interface (touch screen) of the CARDIOHELP device. The human machine interface (HMI) is the central display of the CARDIOHELP device, where measured values, settings and alarms are displayed and adjusted/confirmed, with the exception of bl

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AT · Austria

Affected area 4

AU · Australia

Affected area 5

BE · Belgium

Affected area 6

CA · Canada

Affected area 7

CH · Switzerland

Affected area 8

CZ · Czech Republic

Affected area 9

DE · Germany

Affected area 10

DK · Denmark

Affected area 11

EG · Egypt

Affected area 12

ES · Spain

Affected area 13

FI · Finland

Affected area 14

FR · France

Affected area 15

GB · United Kingdom

Affected area 16

GR · Greece

Affected area 17

HK · Hong Kong

Affected area 18

HR · Croatia

Affected area 19

IE · Ireland

Affected area 20

IS · Iceland

Affected area 21

IT · Italy

Affected area 22

JO · Jordan

Affected area 23

KR · South Korea

Affected area 24

KW · Kuwait

Affected area 25

LU · Luxembourg

Affected area 26

NL · Netherlands

Affected area 27

NO · Norway

Affected area 28

PL · Poland

Affected area 29

PT · Portugal

Affected area 30

RU · Russia

Affected area 31

SA · Saudi Arabia

Affected area 32

SE · Sweden

Affected area 33

SG · Singapore

Affected area 34

SI · Slovenia

Affected area 35

SK · Slovakia

Affected area 36

TH · Thailand

Affected area 37

TR · Turkey

Affected area 38

ZA · South Africa

Company & contacts

Manufacturer

Maquet Cardiovascular, LLC

Phone

Not provided

Brand: Maquet Cardiovascular. Contact: Maquet Cardiovascular, LLC.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1149-2013

Event ID

64641

Units affected

154

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 24 Apr 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button