VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of instruments, software

Official notice

27 Sept 2017

RecallScope risk estimate:MediumFDA

An official FDA notice concerns VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of instruments, software. Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC¿ 700327(TM) strain is positive. Published 2017-09-27.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

11,703

Recall number

Z-3134-2017

Official notice

2017-09-27

The hazard

Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC¿ 700327(TM) strain is positive.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

2420240403, 2420303403

Event Ids

77935

Recall Numbers

Z-3134-2017

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

VITEK(R) 2 GP, REF 21342, IVD, UDI 03573026131920, bioMerieux, Inc. VITEK(R) 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria

Units affected

11,703

Hazard

Customer reports have indicated occurrences of atypical negative O129R reactions for Enterococcus casseliflavus ATCC(R) 700327(TM) in association with VITEK(R) 2 GP ID Lot 2420240403 and 2420303403, causing failure of the Quality Control (QC) testing for the VITEK(R) 2 GP ID card. The expected O129R reaction for ATCC¿ 700327(TM) strain is positive.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AR · Argentina

Affected area 4

AT · Austria

Affected area 5

BE · Belgium

Affected area 6

CA · Canada

Affected area 7

CH · Switzerland

Affected area 8

CN · China

Affected area 9

DE · Germany

Affected area 10

EG · Egypt

Affected area 11

ES · Spain

Affected area 12

FR · France

Affected area 13

GB · United Kingdom

Affected area 14

GR · Greece

Affected area 15

HR · Croatia

Affected area 16

HU · Hungary

Affected area 17

IL · Israel

Affected area 18

IT · Italy

Affected area 19

KR · South Korea

Affected area 20

LU · Luxembourg

Affected area 21

LV · Latvia

Affected area 22

MX · Mexico

Affected area 23

NL · Netherlands

Affected area 24

PL · Poland

Affected area 25

RS · Serbia

Affected area 26

SA · Saudi Arabia

Affected area 27

TR · Turkey

Affected area 28

ZA · South Africa

Company & contacts

Manufacturer

Biomerieux Inc

Phone

Not provided

Brand: Biomerieux. Contact: Biomerieux Inc.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-3134-2017

Event ID

77935

Units affected

11,703

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 27 Sept 2017. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button