Open official notice

NucliSENS Lysis Buffer Extended

Official notice

15 Nov 2017

RecallScope risk estimate:MediumFDA

An official FDA notice concerns NucliSENS Lysis Buffer Extended. Elute may become colored due to the residual presence of heme group origination from hemoglobin present in whole blood samples as well as dried spot (DBS) samples. This can potentially cause the inhibition of PCR reactions and result in an uninterpretable test result. If using an Internal Control (IC) in the extraction process the IC could potentially be inhibited as well and invalidate test results. Published 2017-11-15.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

911

Recall number

Z-0086-2018

Official notice

2017-11-15

The hazard

Elute may become colored due to the residual presence of heme group origination from hemoglobin present in whole blood samples as well as dried spot (DBS) samples. This can potentially cause the inhibition of PCR reactions and result in an uninterpretable test result. If using an Internal Control (IC) in the extraction process the IC could potentially be inhibited as well and invalidate test results.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

17022802

Event Ids

78387

Recall Numbers

Z-0086-2018

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

NucliSENS Lysis Buffer Extended lot for RES 76675

Units affected

911

Hazard

Elute may become colored due to the residual presence of heme group origination from hemoglobin present in whole blood samples as well as dried spot (DBS) samples. This can potentially cause the inhibition of PCR reactions and result in an uninterpretable test result. If using an Internal Control (IC) in the extraction process the IC could potentially be inhibited as well and invalidate test results.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States · CO

Location 2

US · United States · MD

Location 3

US · United States · OH

Affected area 1

US · United States

Company & contacts

Manufacturer

BioMerieux SA

Phone

Not provided

Brand: Biomerieux. Contact: BioMerieux SA.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0086-2018

Event ID

78387

Units affected

911

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 15 Nov 2017. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.