Open official notice

Veritas advanced fluidics and infusion pack

Official notice

15 Jun 2023

RecallScope risk estimate:MediumHealth Canada

An official Health Canada notice concerns Veritas advanced fluidics and infusion pack. Johnson & Johnson Surgical Vision, Inc. Is initiating this new recall due to a manufacturing issue with Veritas packs which could result in a weld protrusion, which is the physical gap between the housing and cover of the Veritas packs, that exceeds the design specification. Recall start date: June 2, 2023 Published 2023-06-15.

What to do now

Contact the manufacturer if you require additional information.

Recall number

73826

Official notice

2023-06-15

The hazard

Johnson & Johnson Surgical Vision, Inc. Is initiating this new recall due to a manufacturing issue with Veritas packs which could result in a weld protrusion, which is the physical gap between the housing and cover of the Veritas packs, that exceeds the design specification. Recall start date: June 2, 2023

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

73826

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - Veritas Advanced Fluidics Pack, Lot or serial number - All lots., Model or catalog number - VRT-AF

Product 2

Affected products - Veritas Advanced Infusion Pack, Lot or serial number - All lots., Model or catalog number - VRT-AI

Hazard

Johnson & Johnson Surgical Vision, Inc. Is initiating this new recall due to a manufacturing issue with Veritas packs which could result in a weld protrusion, which is the physical gap between the housing and cover of the Veritas packs, that exceeds the design specification. Recall start date: June 2, 2023

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Johnson Johnson Surgical Vision, Veritas Advanced Fluidics And Infusion Pack.

Source & status

Tier

official

Category

Health

Official classification

Type II

Recall details

Recall number

73826

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 15 Jun 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button