Veritas advanced infusion pack

Official notice

10 Oct 2022

RecallScope risk estimate:CriticalHealth Canada

An official Health Canada notice concerns Veritas advanced infusion pack. Johnson & Johnson vision has initiated this action due to the potential for the irrigation luer to crack or break. A crack or break of the irrigation luer could lead to reduced irrigation pressure during surgery, associated with an unstable anterior chamber. This hazard could result in patient harm such as, but not limited to, capsule tear. Additionally, there could be a delay in surgery if a leakage is noticed during or prior to surgery and the pack must be switched out. As of August 29, 2022, there have been forty-one (41) product complaints, including one (1) adverse event, associated with this issue. Recall start date: September 7, 2022 Published 2022-10-10.

What to do now

Contact manufacturer if you require additional information.

Recall number

72017

Official notice

2022-10-10

The hazard

Johnson & Johnson vision has initiated this action due to the potential for the irrigation luer to crack or break. A crack or break of the irrigation luer could lead to reduced irrigation pressure during surgery, associated with an unstable anterior chamber. This hazard could result in patient harm such as, but not limited to, capsule tear. Additionally, there could be a delay in surgery if a leakage is noticed during or prior to surgery and the pack must be switched out. As of August 29, 2022, there have been forty-one (41) product complaints, including one (1) adverse event, associated with this issue. Recall start date: September 7, 2022

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Recall Numbers

72017

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Affected products - Veritas Advanced Infusion Pack, Lot or serial number - 60330206, Model or catalogue number - VRT-AI

Hazard

Johnson & Johnson vision has initiated this action due to the potential for the irrigation luer to crack or break. A crack or break of the irrigation luer could lead to reduced irrigation pressure during surgery, associated with an unstable anterior chamber. This hazard could result in patient harm such as, but not limited to, capsule tear. Additionally, there could be a delay in surgery if a leakage is noticed during or prior to surgery and the pack must be switched out. As of August 29, 2022, there have been forty-one (41) product complaints, including one (1) adverse event, associated with this issue. Recall start date: September 7, 2022

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Johnson Johnson Surgical Vision, Veritas Advanced Infusion Pack.

Source & status

Tier

official

Category

Health

Official classification

Type II

Recall details

Recall number

72017

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 10 Oct 2022. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button