Varian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. Product Usage: The

Official notice

22 May 2013

RecallScope risk estimate:MediumFDA

Varian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. Product Usage: The Varian High Energy Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

3,943

Recall number

Z-1325-2013

Official notice

2013-05-22

The hazard

Under certain conditions, photon beams in High Energy Clinacs may experience a gradual change in beam symmetry, potentially reaching an asymmetry of up to about 7% before interlock occurs. At maximum asymmetry, this may result in no more than about 3.5% dose deviation from expected at any point in the beam, and may result in minor injury to the patient or slightly decreased local disease control.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Event Ids

64997

Recall Numbers

Z-1325-2013

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Varian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29, HCX. Product Usage: The Varian High Energy Accelerator is intended to provide stereotactic radiosurgery and precision radiotherapy for lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

Units affected

3,943

Hazard

Under certain conditions, photon beams in High Energy Clinacs may experience a gradual change in beam symmetry, potentially reaching an asymmetry of up to about 7% before interlock occurs. At maximum asymmetry, this may result in no more than about 3.5% dose deviation from expected at any point in the beam, and may result in minor injury to the patient or slightly decreased local disease control.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

AE · United Arab Emirates

Affected area 3

AT · Austria

Affected area 4

AU · Australia

Affected area 5

BG · Bulgaria

Affected area 6

BR · Brazil

Affected area 7

CA · Canada

Affected area 8

CH · Switzerland

Affected area 9

CL · Chile

Affected area 10

CN · China

Affected area 11

CY · Cyprus

Affected area 12

DE · Germany

Affected area 13

DK · Denmark

Affected area 14

DO · Dominican Republic

Affected area 15

EE · Estonia

Affected area 16

EG · Egypt

Affected area 17

ES · Spain

Affected area 18

FI · Finland

Affected area 19

GB · United Kingdom

Affected area 20

GR · Greece

Affected area 21

HK · Hong Kong

Affected area 22

IE · Ireland

Affected area 23

IL · Israel

Affected area 24

IN · India

Affected area 25

IT · Italy

Affected area 26

JP · Japan

Affected area 27

KR · South Korea

Affected area 28

KW · Kuwait

Affected area 29

LT · Lithuania

Affected area 30

LV · Latvia

Affected area 31

MO · Macau

Affected area 32

MX · Mexico

Affected area 33

MY · Malaysia

Affected area 34

NL · Netherlands

Affected area 35

NO · Norway

Affected area 36

NZ · New Zealand

Affected area 37

PE · Peru

Affected area 38

PK · Pakistan

Affected area 39

PL · Poland

Affected area 40

PT · Portugal

Affected area 41

RO · Romania

Affected area 42

RS · Serbia

Affected area 43

RU · Russia

Affected area 44

SE · Sweden

Affected area 45

SG · Singapore

Affected area 46

SK · Slovakia

Affected area 47

TH · Thailand

Affected area 48

TR · Turkey

Affected area 49

TW · Taiwan

Affected area 50

UA · Ukraine

Affected area 51

UY · Uruguay

Affected area 52

VN · Vietnam

Affected area 53

ZA · South Africa

Company & contacts

Manufacturer

Varian Medical Systems, Inc. Oncology Systems

Phone

Not provided

Brand: Varian Medical Oncology. Contact: Varian Medical Systems, Inc. Oncology Systems.

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1325-2013

Event ID

64997

Units affected

3,943

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 22 May 2013. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button