Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811
Official notice
11 Nov 2026
Last checked
17 Nov 2026
An official FDA notice concerns Tandem Mobi Insulin Pump, Models: 1004004, 1010750, 1012676, 1012719, 1013655, 1013656, 1013501, 1013700, 1014081, 1014801, 1014811. Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization. Published 2026-11-11.
What to do now
Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.
Units affected
80,938
Recall number
Z-0316-2027
Official notice
2026-11-11
The hazard
The hazard
Tandem Diabetes Care, Inc. ( Tandem ) has identified that certain Tandem Mobi insulin pumps may generate false-positive occlusion alarms when no actual insulin flow obstruction is present. Investigation determined that increased friction within the drive module can reduce available motor power and, under certain conditions, cause motor speed to fall below the occlusion detection threshold. As a result, the pump may incorrectly detect an occlusion condition and suspend insulin delivery. When an occlusion alarm occurs, insulin delivery remains suspended until the user acknowledges the alarm and resumes therapy. Repeated false-positive occlusion alarms, if not addressed promptly, may result in interruptions of insulin delivery that can increase the risk of elevated blood glucose levels. In severe cases, hyperglycemia may require medical intervention or hospitalization.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Other
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Affected area 2
Company & contacts
Company & contacts
Brand: Tandem Diabetes Care. Contact: Tandem Diabetes Care, Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.