Image not available in RecallScope.
Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference FSCA Identifier: CP-08881
Official notice
26 Sept 2012
An official FDA notice concerns Varian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29, Reference FSCA Identifier: CP-08881. Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-dose of the intended target volume. Published 2012-09-26.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
221
Recall number
Z-2383-2012
Official notice
2012-09-26
The hazard
The hazard
Some X-jaw (lower collimator jaw) carriers were made using the incorrect metal alloy, which can cause the jaw carrier to crack. If both carriers on the same X-jaw crack, the jaw would be able to move freely in the closed direction. The position readout interlock circuitry will not detect this jaw position deviation. This may lead to treatment with an under-dose of the intended target volume.
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Manufacturer
Phone
Brand: Varian Medical Oncology. Contact: Varian Medical Systems, Inc. Oncology Systems.
Source & status
Source & status
Recall details
Recall details
Recall number
Event id
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.