No image published for this recall

No image published for this recall

Trusystem 7500 surgical table

Official notice

7 Mar 2025

Last checked

7 Mar 2025

Medium riskHealth Canada

An official Health Canada notice concerns Trusystem 7500 surgical table. Baxter Corporation is issuing an urgent medical device correction for the TruSystem 7500 Surgical Table Systems due to a software issue which causes the upper back section of model 26-series tabletops to become inoperable (non-adjustable) when the emergency mode function is enabled. Recall Start Date: February 27, 2025 Published 2025-03-07.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77080

Official notice

2025-03-07

The hazard

Baxter Corporation is issuing an urgent medical device correction for the TruSystem 7500 Surgical Table Systems due to a software issue which causes the upper back section of model 26-series tabletops to become inoperable (non-adjustable) when the emergency mode function is enabled. Recall Start Date: February 27, 2025

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

77080

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - TruSystem 7500 Surgical Table, Lot or serial number - 109083213 108232406 102905380 108618965 102065305 108574824 107616874, Model or catalog number - 1704695 1730732 1854087 4091000

Risk / hazard

Baxter Corporation is issuing an urgent medical device correction for the TruSystem 7500 Surgical Table Systems due to a software issue which causes the upper back section of model 26-series tabletops to become inoperable (non-adjustable) when the emergency mode function is enabled. Recall Start Date: February 27, 2025

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Baxter Medical T Kg, Trusystem 7500 Surgical Table.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type II

App estimate

medium

Recall details

Recall number

77080

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 7 Mar 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.