Pst 500 u precision surgical table

Official notice

4 Jul 2024

Last checked

4 Jul 2024

Critical riskHealth Canada

An official Health Canada notice concerns Pst 500 u precision surgical table. Baxter Corporation is issuing an urgent medical device correction for the device due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport. A sudden drop and/or tilt of the surgical tabletop during a surgical procedure may cause unintended patient movement or fall. This could lead to delay or interruption in therapy, ineffective CPR, and could cause critical patient harm such as major musculoskeletal or vital organ injuries. Baxter has received six complaints related to this issue; however, no serious injury was reported. Baxter will correct this issue by providing a service kit that can be installed either by a hospital technician or by a Baxter technician or authorized representative. If the service kit is self-installed, a Baxter technician or authorized representative will follow up to verify proper installation. Recall start date: July 2, 2024 Published 2024-07-04.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75784

Official notice

2024-07-04

The hazard

Baxter Corporation is issuing an urgent medical device correction for the device due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport. A sudden drop and/or tilt of the surgical tabletop during a surgical procedure may cause unintended patient movement or fall. This could lead to delay or interruption in therapy, ineffective CPR, and could cause critical patient harm such as major musculoskeletal or vital organ injuries. Baxter has received six complaints related to this issue; however, no serious injury was reported. Baxter will correct this issue by providing a service kit that can be installed either by a hospital technician or by a Baxter technician or authorized representative. If the service kit is self-installed, a Baxter technician or authorized representative will follow up to verify proper installation. Recall start date: July 2, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75784

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - PST 500 U Precision Surgical Table, Lot or serial number - 109148151 108029053 108167723 108649034 108664873 109393980 109395301 108167561 108424302, Model or catalog number - 4080300

Risk / hazard

Baxter Corporation is issuing an urgent medical device correction for the device due to a potential issue with loose spring pins in the spindle drives. This can cause the tabletop to unexpectedly tilt or move at any time, even without active use of the surgical table. This could potentially result in unintentional patient movement during surgical procedures and/or preparation for surgical procedures, including transport. A sudden drop and/or tilt of the surgical tabletop during a surgical procedure may cause unintended patient movement or fall. This could lead to delay or interruption in therapy, ineffective CPR, and could cause critical patient harm such as major musculoskeletal or vital organ injuries. Baxter has received six complaints related to this issue; however, no serious injury was reported. Baxter will correct this issue by providing a service kit that can be installed either by a hospital technician or by a Baxter technician or authorized representative. If the service kit is self-installed, a Baxter technician or authorized representative will follow up to verify proper installation. Recall start date: July 2, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Baxter Medical T Kg, Pst 500 U Precision Surgical Table.

Source & status

Source

Health Canada

Tier

official

Category

HOME

Classification

Type I

App estimate

critical

Recall details

Recall number

75784

Classification

Type I

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 4 Jul 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.