No image published for this recall
An official Health Canada notice concerns Somatom force. During the transmission of the patient table position information, sporadic interference signals may occur due to an error within the patient table position encoder. This problem can be detected by the system and can result in a scan abort. Sporadic aborts of perfusion scans related to this issue have been reported to us, which can occur during the patient table's back and forth movement when using the adaptive 4D spiral. This update instruction provides the revision 01 of the position sensor for the table top of affected PHS-5 and MPT-4 patient tables to prevent sporadic signal glitches at the zigzag movement of perfusion scans. Recall start date: Apr 28, 2022 Published 2022-05-03.
What to do now
Contact the manufacturer if you require additional information.
Recall number
64143
Official notice
2022-05-03
The hazard
The hazard
During the transmission of the patient table position information, sporadic interference signals may occur due to an error within the patient table position encoder. This problem can be detected by the system and can result in a scan abort. Sporadic aborts of perfusion scans related to this issue have been reported to us, which can occur during the patient table's back and forth movement when using the adaptive 4D spiral. This update instruction provides the revision 01 of the position sensor for the table top of affected PHS-5 and MPT-4 patient tables to prevent sporadic signal glitches at the zigzag movement of perfusion scans. Recall start date: Apr 28, 2022
How to check if you're affected
How to check if you're affected
Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.
Recall Numbers
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected vehicles were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand
Manufacturer
Phone
Brand: Siemens, Somatom Force.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Artis icono ceiling

Artis icono ceiling, floor and biplane systems

Magnetom prisma, vida, skyra, verio systems

Artis pheno

Uroskop omnia

Atellica ch reaction ring cuvette segment

Advia® and atellica® chemistry urinary cerebrospinal fluid protein (ucfp) assay

Luminos lotus max, drf max, and agile max

Luminos drf max, mobilett elara max, ysio max and ysio x.pree

Axiom vertix and axiom multix

Siemens biograph mmr system

Artis icono ceiling
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 3 May 2022. The next one won't wait for you to search.