An official Health Canada notice concerns Artis pheno. Initiating a shutdown of the system can sporadically lead to a state in which the robotic C-arm stand cannot be moved after the next system startup. Other movements (e.g. Table movements) are still available with reduced speed. The system may show the following generic messages "reduced stand/table speed" and "limited stand movement, wait". In case the error occurs the C-arm stand will not move anymore. Therefore, the C-arm cannot be moved into the required position. This may result in a situation where it is necessary to cancel the clinical procedure or to begin or continue the procedure on an alternative system. Recall start date: August 13, 2025 Published 2025-08-20.
What to do now
Contact the manufacturer if you require additional information.
Recall number
77943
Official notice
2025-08-20
The hazard
The hazard
Initiating a shutdown of the system can sporadically lead to a state in which the robotic C-arm stand cannot be moved after the next system startup. Other movements (e.g. Table movements) are still available with reduced speed. The system may show the following generic messages "reduced stand/table speed" and "limited stand movement, wait". In case the error occurs the C-arm stand will not move anymore. Therefore, the C-arm cannot be moved into the required position. This may result in a situation where it is necessary to cancel the clinical procedure or to begin or continue the procedure on an alternative system. Recall start date: August 13, 2025
How to check if you're affected
How to check if you're affected
Compare the product name and identifiers below with the item you own. Follow the official notice if they match.
Recall Numbers
Checked manually this time?
The app alerts you when a new official recall is published — no searching, no checking back.
Turn on recall alertsAffected products
Affected products
Product 1
Product 2
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected products were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Siemens, Artis Pheno.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Artis icono ceiling, floor and biplane systems
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Artis icono ceiling
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Magnetom prisma, vida, skyra, verio systems
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Uroskop omnia
Brand match
Critical risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Advia® and atellica® chemistry urinary cerebrospinal fluid protein (ucfp) assay
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Axiom vertix and axiom multix
Brand match
High risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Siemens biograph mmr system
Brand match
Critical risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Artis icono ceiling
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Cios spin, cios flow, and cios alpha
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Magnetom mri system
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Mobilett mira max
Brand match
Medium risk
Markets
No image published for this recall
The official notice did not include a photo. Check the product details and identifiers below.
Mammomat revelation
Brand match
Medium risk
Markets
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 20 Aug 2025. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.