No image published for this recall
Advia® and atellica® chemistry urinary cerebrospinal fluid protein (ucfp) assay
Official notice
21 Aug 2024
Last checked
21 Aug 2024
An official Health Canada notice concerns Advia® and atellica® chemistry urinary cerebrospinal fluid protein (ucfp) assay. Siemens Healthineers has confirmed the potential for biased quality control (qc) and patient results when using Atellica® CH Urinary/Cerebrospinal Fluid Protein (UCFP) lot 130414 on the Stellica® CH and Stellica® CI analyzers. Siemens' internal investigation confirmed when using lot 130414, QC may recover outside of the allowable control limits for urine chemistry and spinal fluid control levels. Recall start date: August 1, 2024 Published 2024-08-21.
What to do now
Contact the manufacturer if you require additional information.
Recall number
75952
Official notice
2024-08-21
The hazard
The hazard
Siemens Healthineers has confirmed the potential for biased quality control (qc) and patient results when using Atellica® CH Urinary/Cerebrospinal Fluid Protein (UCFP) lot 130414 on the Stellica® CH and Stellica® CI analyzers. Siemens' internal investigation confirmed when using lot 130414, QC may recover outside of the allowable control limits for urine chemistry and spinal fluid control levels. Recall start date: August 1, 2024
How to check if you're affected
How to check if you're affected
Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.
Lots
Recall Numbers
Checked manually this time?
The app watches your VIN and alerts you automatically next time — no searching, no checking back.
Turn on automatic VIN checksAffected products
Affected products
Product 1
Product 2
Risk / hazard
Where it was sold
Where it was sold
The source does not list specific retailers. See the official notice for where affected vehicles were sold.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand
Manufacturer
Phone
Brand: Siemens, Advia® And Atellica® Chemistry Urinary Cerebrospinal Fluid Protein Assay.
Source & status
Source & status
Source
Tier
Category
Classification
App estimate
Recall details
Recall details
Recall number
Classification
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeRelated recalls

Artis icono ceiling

Artis icono ceiling, floor and biplane systems

Magnetom prisma, vida, skyra, verio systems

Artis pheno

Uroskop omnia

Atellica ch reaction ring cuvette segment

Luminos lotus max, drf max, and agile max

Luminos drf max, mobilett elara max, ysio max and ysio x.pree

Axiom vertix and axiom multix

Siemens biograph mmr system

Artis icono ceiling

Cios spin, cios flow, and cios alpha
Weekly recall digest
Get next week's recalls before you need them.
One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.
This recall was published 21 Aug 2024. The next one won't wait for you to search.
RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.