Pro+ mattress

Official notice

11 Jul 2022

MediumHealth Canada

An official Health Canada notice concerns Pro+ mattress. The Pro+ Mattress with Advanced Microclimate® Technology (AMT) uses a non-powered system of intake and output valves to provide weight-based pressure redistribution whenever a patient moves or is repositioned. The Pro+ hospital bed surface is a mattress used with various Hillrom hospital beds such as Centrella, Versa care, Advanta, Careassist and Hillrom900 beds. A total of 139 complaints from one customer were received for Centrella bed and mattress issues with Low Air Loss (LAL) errors on the mcm display system from 1 Jan 2019 (earliest distribution date) through 1 March 2022. Initially complaints were raised against the beds, however investigation determined the issue was with the Pro+ Mattress blower system. On 9 March 2022, during complaint investigation, service technicians troubleshooted the issue by reconnecting the system to fix the alarm and error. Log file review later identified several errors in the display which confirmed that the troubleshooting (reconnection of the systems) did not solve the issue. The engineering team determined that the error messages were caused by the blower of the mattress with Low Air Loss (LAL). The impact of this failure mode can cause a reduction in the microclimate management performance of the Pro+, which could result in an increased risk to skin and result in potential pressure injury. Microclimate management is used to reduce skin temperature and moisture of patient skin. No adverse patient experiences have been reported. Recall start date: July 1, 2022 Published 2022-07-11.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64351

Official notice

2022-07-11

The hazard

The Pro+ Mattress with Advanced Microclimate® Technology (AMT) uses a non-powered system of intake and output valves to provide weight-based pressure redistribution whenever a patient moves or is repositioned. The Pro+ hospital bed surface is a mattress used with various Hillrom hospital beds such as Centrella, Versa care, Advanta, Careassist and Hillrom900 beds. A total of 139 complaints from one customer were received for Centrella bed and mattress issues with Low Air Loss (LAL) errors on the mcm display system from 1 Jan 2019 (earliest distribution date) through 1 March 2022. Initially complaints were raised against the beds, however investigation determined the issue was with the Pro+ Mattress blower system. On 9 March 2022, during complaint investigation, service technicians troubleshooted the issue by reconnecting the system to fix the alarm and error. Log file review later identified several errors in the display which confirmed that the troubleshooting (reconnection of the systems) did not solve the issue. The engineering team determined that the error messages were caused by the blower of the mattress with Low Air Loss (LAL). The impact of this failure mode can cause a reduction in the microclimate management performance of the Pro+, which could result in an increased risk to skin and result in potential pressure injury. Microclimate management is used to reduce skin temperature and moisture of patient skin. No adverse patient experiences have been reported. Recall start date: July 1, 2022

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

64351

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Pro+ Mattress, Lot or serial number - Not applicable., Model or catalog number - P7924A03

Product 2

Affected products - Pro+ Mattress, Lot or serial number - Not applicable., Model or catalog number - P7923A04 P7923A03

Product 3

Affected products - Pro+ Mattress, Lot or serial number - Not applicable., Model or catalog number - P7924A01

Product 4

Affected products - Pro+ Mattress, Lot or serial number - Not applicable., Model or catalog number - P3255A01

Product 5

Affected products - Pro+ Mattress, Lot or serial number - Not applicable., Model or catalog number - P7923A01

Product 6

Affected products - Pro+ Mattress, Lot or serial number - Not applicable., Model or catalog number - P3255A02

Risk / hazard

The Pro+ Mattress with Advanced Microclimate® Technology (AMT) uses a non-powered system of intake and output valves to provide weight-based pressure redistribution whenever a patient moves or is repositioned. The Pro+ hospital bed surface is a mattress used with various Hillrom hospital beds such as Centrella, Versa care, Advanta, Careassist and Hillrom900 beds. A total of 139 complaints from one customer were received for Centrella bed and mattress issues with Low Air Loss (LAL) errors on the mcm display system from 1 Jan 2019 (earliest distribution date) through 1 March 2022. Initially complaints were raised against the beds, however investigation determined the issue was with the Pro+ Mattress blower system. On 9 March 2022, during complaint investigation, service technicians troubleshooted the issue by reconnecting the system to fix the alarm and error. Log file review later identified several errors in the display which confirmed that the troubleshooting (reconnection of the systems) did not solve the issue. The engineering team determined that the error messages were caused by the blower of the mattress with Low Air Loss (LAL). The impact of this failure mode can cause a reduction in the microclimate management performance of the Pro+, which could result in an increased risk to skin and result in potential pressure injury. Microclimate management is used to reduce skin temperature and moisture of patient skin. No adverse patient experiences have been reported. Recall start date: July 1, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: Hill-rom, Pro+ Mattress.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

64351

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

HEALTH
Critical

Volara system patient ciruit

Health Canada29 May 2026

Brand match

Critical risk

Markets

CA
HEALTH
Medium

Centrella smart+ bed

Health Canada5 Aug 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Affinity 4 birthing bed

Health Canada22 May 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Centrella max and pro+ mattress replacement system (mrs) hospital bed mattresses

Health Canada2 Jul 2024

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Volara system single patient use circuit

Health Canada2 Jul 2024

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Navicare patient safety; voalte patient safety

Health Canada2 Jun 2023

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Pro+ mattress

Health Canada13 Apr 2023

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Progressa (frame)

Health Canada10 Nov 2022

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Navicare nurse call

Health Canada7 Jun 2022

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Navicare nurse call (software)

Health Canada26 Nov 2021

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Dlp select 3d arterial cannula and dlp elongated one piece arterial cannulae

Health Canada3 Jul 2026

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Intramedullary bone saw blade assembly

Health Canada3 Jul 2026

Brand match

Medium risk

Markets

CA

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 11 Jul 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.