Centrella max and pro+ mattress replacement system (mrs) hospital bed mattresses

Official notice

2 Jul 2024

MediumHealth Canada

An official Health Canada notice concerns Centrella max and pro+ mattress replacement system (mrs) hospital bed mattresses. Baxter Corporation is issuing a correction for the Centrella Max and Pro+ Mattress Replacement System (MRS) Hospital Bed Mattresses due to the potential for the top cover to delaminate. Contributing causes of delamination include improper cleaning of the product (i.e., excessive use of cleaning agents containing bleach) and a lack of adequate details on the appropriate use of cleaning agents containing bleach in the product instructions for use (IFU). Recall Start Date: May 13, 2024 Published 2024-07-02.

What to do now

Contact the manufacturer if you require additional information.

Recall number

75768

Official notice

2024-07-02

The hazard

Baxter Corporation is issuing a correction for the Centrella Max and Pro+ Mattress Replacement System (MRS) Hospital Bed Mattresses due to the potential for the top cover to delaminate. Contributing causes of delamination include improper cleaning of the product (i.e., excessive use of cleaning agents containing bleach) and a lack of adequate details on the appropriate use of cleaning agents containing bleach in the product instructions for use (IFU). Recall Start Date: May 13, 2024

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Recall Numbers

75768

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Affected products - Pro+ MRS Surface, Lot or serial number - All lots distributed from March 1, 2022., Model or catalog number - P7924A03 P7924A04

Product 2

Affected products - Pro+ MRS Surface For Hillrom 900 With X-Ray, Lot or serial number - All lots distributed from March 1, 2022., Model or catalog number - P006800A01

Product 3

Affected products - Centrella Pro+ Surface With X-Ray, Lot or serial number - All produced after June 9, 2020., Model or catalog number - P7923A02 P7923A01

Product 4

Affected products - Pro+ MRS Surface For VersaCare, Lot or serial number - All lots distributed from March 1, 2022., Model or catalog number - P3255A02

Product 5

Affected products - Pro+ MRS Surface With X-Ray, Lot or serial number - All produced after June 9, 2020., Model or catalog number - P7924A02 P7924A01

Product 6

Affected products - Centrella Max Surface, Lot or serial number - All produced after June 9, 2020., Model or catalog number - P7922A01 P7922A02

Product 7

Affected products - Centrella Pro+ Surface, Lot or serial number - All lots distributed from March 1, 2022., Model or catalog number - P7923A04 P7923A03

Product 8

Affected products - Pro+ MRS Surface For VersaCare With X-Ray, Lot or serial number - All produced after June 9, 2020., Model or catalog number - P3255A01

Risk / hazard

Baxter Corporation is issuing a correction for the Centrella Max and Pro+ Mattress Replacement System (MRS) Hospital Bed Mattresses due to the potential for the top cover to delaminate. Contributing causes of delamination include improper cleaning of the product (i.e., excessive use of cleaning agents containing bleach) and a lack of adequate details on the appropriate use of cleaning agents containing bleach in the product instructions for use (IFU). Recall Start Date: May 13, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand: Hill-rom, Centrella Max And Pro+ Mattress Replacement System Hospital Bed Mattresses.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

75768

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

HEALTH
Critical

Volara system patient ciruit

Health Canada29 May 2026

Brand match

Critical risk

Markets

CA
HEALTH
Medium

Centrella smart+ bed

Health Canada5 Aug 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Affinity 4 birthing bed

Health Canada22 May 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Volara system single patient use circuit

Health Canada2 Jul 2024

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Navicare patient safety; voalte patient safety

Health Canada2 Jun 2023

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Pro+ mattress

Health Canada13 Apr 2023

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Pro+ mattress

Health Canada11 Jul 2022

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Navicare nurse call

Health Canada7 Jun 2022

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Navicare nurse call (software)

Health Canada26 Nov 2021

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Progressa (frame)

Health Canada10 Nov 2022

Brand match

Medium risk

Markets

CA
HEALTH
High

Abiomed introducer kits with impella heart pumps

Health Canada29 Jun 2026

Brand match

High risk

Markets

CA
HEALTH
Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)Medium

Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Health Canada22 Jun 2026

Brand match

Medium risk

Markets

CA

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 2 Jul 2024. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.