Recall image placeholder

No image published for this recall

Centrella smart+ bed

Official notice

5 Aug 2025

Last checked

5 Aug 2025

Medium riskHealth Canada
Watch Hill-rom

An official Health Canada notice concerns Centrella smart+ bed. Baxter Corporation is issuing an urgent medical device correction for Centrella Smart+ beds with system software version 1.27.000 due to the potential for an interface issue with the Rauland Responder 5 nurse call system. The bed exit system will alert at the bed but may fail to send a remote alert through the nurse call system if a remote alert was previously sent and cancelled at the in-room nurse call wall unit. If the bed exit alert fails to notify the caregiver at the assigned remote location (e.g., nurse's station) the staff may be unaware the patient exited the bed if they are unable to hear the local bed alert. As a result, a patient may experience a fall resulting in a critical musculoskeletal injury, including head trauma. To date, Baxter has not received any reports of patient injury related to this issue. Recall start date: September 19, 2024 Published 2025-08-05.

What to do now

Contact the manufacturer if you require additional information.

Recall number

77811

Official notice

2025-08-05

The hazard

Baxter Corporation is issuing an urgent medical device correction for Centrella Smart+ beds with system software version 1.27.000 due to the potential for an interface issue with the Rauland Responder 5 nurse call system. The bed exit system will alert at the bed but may fail to send a remote alert through the nurse call system if a remote alert was previously sent and cancelled at the in-room nurse call wall unit. If the bed exit alert fails to notify the caregiver at the assigned remote location (e.g., nurse's station) the staff may be unaware the patient exited the bed if they are unable to hear the local bed alert. As a result, a patient may experience a fall resulting in a critical musculoskeletal injury, including head trauma. To date, Baxter has not received any reports of patient injury related to this issue. Recall start date: September 19, 2024

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Recall Numbers

77811

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected products - Centrella Smart+ Bed, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - P7900B100205

Product 2

Affected products - Centrella Smart+ Bed, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - P7900B100021

Product 3

Affected products - Centrella Smart+ Bed, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - P7900B100020

Product 4

Affected products - Centrella Smart+ Bed, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - P7900B100019

Product 5

Affected products - Centrella Smart+ Bed, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - P7900B100223

Product 6

Affected products - Centrella Smart+ Bed, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - P7900B100224

Product 7

Affected products - Centrella Smart+ Bed, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - P7900B100231

Risk / hazard

Baxter Corporation is issuing an urgent medical device correction for Centrella Smart+ beds with system software version 1.27.000 due to the potential for an interface issue with the Rauland Responder 5 nurse call system. The bed exit system will alert at the bed but may fail to send a remote alert through the nurse call system if a remote alert was previously sent and cancelled at the in-room nurse call wall unit. If the bed exit alert fails to notify the caregiver at the assigned remote location (e.g., nurse's station) the staff may be unaware the patient exited the bed if they are unable to hear the local bed alert. As a result, a patient may experience a fall resulting in a critical musculoskeletal injury, including head trauma. To date, Baxter has not received any reports of patient injury related to this issue. Recall start date: September 19, 2024

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Hill-rom, Smart

Manufacturer

Not provided

Phone

Not provided

Brand: Hill-rom, Smart.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

77811

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Volara system patient ciruit

Critical riskHealth Canada29 May 2026
CA
Recall image placeholder

Affinity 4 birthing bed

Medium riskHealth Canada22 May 2025
CA
Recall image placeholder

Continuous glucose monitoring smart device software applications

Risk not stated by sourceHealth Canada19 Mar 2025
CA
Recall image placeholder

Centrella max and pro+ mattress replacement system (mrs) hospital bed mattresses

Medium riskHealth Canada2 Jul 2024
CA
Recall image placeholder

Volara system single patient use circuit

Medium riskHealth Canada2 Jul 2024
CA
Recall image placeholder

Navicare patient safety; voalte patient safety

Medium riskHealth Canada2 Jun 2023
CA
Recall image placeholder

Pro+ mattress

Medium riskHealth Canada13 Apr 2023
CA
Recall image placeholder

Progressa (frame)

Medium riskHealth Canada10 Nov 2022
CA
Recall image placeholder

Pro+ mattress

Medium riskHealth Canada11 Jul 2022
CA
Recall image placeholder

Navicare nurse call

Medium riskHealth Canada7 Jun 2022
CA
Recall image placeholder

Smart toe ii intramedullary arthrodesis implant

Medium riskHealth Canada11 Apr 2022
CA
Recall image placeholder

Navicare nurse call (software)

Medium riskHealth Canada26 Nov 2021

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 5 Aug 2025. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.