Philips intellivue patient monitors

Official notice

20 Feb 2026

Last checked

20 Feb 2026

Critical riskHealth Canada
Watch Philips

An official Health Canada notice concerns Philips intellivue patient monitors. Philips IntelliVue patient monitors provide a feature for pausing all alarms for a configured timeframe or for switching all alarms off for an infinite period until they are switched back on by the user. The infinite alarms off feature is an important functionality in specific clinical environments such as the operating room, cath lab or during resuscitations to reduce the audible alarms which may present unnecessary distractions during closely monitored procedures or interventions. Philips has received complaints that the IntelliVue monitors did not alarm. In some cases, these complaints alleged delayed patient treatment due to the monitors not alarming; and, in some cases, there were reports of severe patient harm or death. It was determined that in these instances the alarms had been switched off for an infinite period, and the user may have been unaware that the alarms were off. Published 2026-02-20.

What to do now

Contact the manufacturer if you require additional information.

Recall number

81234

Official notice

2026-02-20

The hazard

Philips IntelliVue patient monitors provide a feature for pausing all alarms for a configured timeframe or for switching all alarms off for an infinite period until they are switched back on by the user. The infinite alarms off feature is an important functionality in specific clinical environments such as the operating room, cath lab or during resuscitations to reduce the audible alarms which may present unnecessary distractions during closely monitored procedures or interventions. Philips has received complaints that the IntelliVue monitors did not alarm. In some cases, these complaints alleged delayed patient treatment due to the monitors not alarming; and, in some cases, there were reports of severe patient harm or death. It was determined that in these instances the alarms had been switched off for an infinite period, and the user may have been unaware that the alarms were off.

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Skus

M8003A, 866060, 866470, M8007A, MX700, M8005A, M8001A, M8105A, M3002A, M8010A, M8002A, M8008A, M8004A, M8102A, 866066, 866471, 866064, 866062, MX800, 867033, 867030

Other

RA RA-81234

Model Numbers

M8003A, 866060, 866470, M8007A, MX700, M8005A, M8001A, M8105A, M3002A, M8010A, M8002A, M8008A, M8004A, M8102A, 866066, 866471, 866064, 866062, MX800, 867033, 867030

Recall Numbers

81234

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected products - IntelliVue MP40 Patient Monitor System - Main Unit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8003A

Product 2

Affected products - IntelliVue MX400 Patient Monitor, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866060

Product 3

Affected products - IntelliVue MX850 Patient Monitor, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866470

Product 4

Affected products - IntelliVue MP70 Patient Monitor System - Main Unit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8007A

Product 5

Affected products - IntelliVue MX Patient Monitor System - Main Unit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - MX700

Product 6

Affected products - IntelliVue MP60 Patient Monitor System - Main Unit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8005A

Product 7

Affected products - IntelliVue MP20 Patient Monitor System, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8001A

Product 8

Affected products - IntelliVue MP5 Patient Monitor System - Main Unit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8105A

Product 9

Affected products - IntelliVue X2 Multi-Measurement Module - Main Unit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M3002A

Product 10

Affected products - IntelliVue MP90 Patient Monitor System - Main Unit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8010A

Product 11

Affected products - IntelliVue MP30 Patient Monitor System, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8002A

Product 12

Affected products - IntelliVue MP80 Patient Monitor System - Main Unit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8008A

Product 13

Affected products - IntelliVue MP50 Patient Monitor System - Main Unit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8004A

Product 14

Affected products - IntelliVue MP2 Patient Monitor - Main Unit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8102A

Product 15

Affected products - IntelliVue MX550 Patient Monitor, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866066

Product 16

Affected products - IntelliVue MX750 Patient Monitor, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866471

Product 17

Affected products - IntelliVue MX500 Patient Monitor, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866064

Product 18

Affected products - IntelliVue Patient Monitor, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866062

Product 19

Affected products - IntelliVue MX Patient Monitor System - Main Unit, Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - MX800

Product 20

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8003A

Product 21

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866060

Product 22

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866470

Product 23

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8007A

Product 24

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - MX700

Product 25

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8005A

Product 26

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8001A

Product 27

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8105A

Product 28

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M3002A

Product 29

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8010A

Product 30

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8002A

Product 31

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8008A

Product 32

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8004A

Product 33

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - M8102A

Product 34

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866066

Product 35

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866471

Product 36

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866064

Product 37

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 866062

Product 38

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - MX800

Product 39

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 867033

Product 40

Lot or serial number - More than 10 numbers, contact manufacturer., Model or catalogue number - 867030

Risk / hazard

Philips IntelliVue patient monitors provide a feature for pausing all alarms for a configured timeframe or for switching all alarms off for an infinite period until they are switched back on by the user. The infinite alarms off feature is an important functionality in specific clinical environments such as the operating room, cath lab or during resuscitations to reduce the audible alarms which may present unnecessary distractions during closely monitored procedures or interventions. Philips has received complaints that the IntelliVue monitors did not alarm. In some cases, these complaints alleged delayed patient treatment due to the monitors not alarming; and, in some cases, there were reports of severe patient harm or death. It was determined that in these instances the alarms had been switched off for an infinite period, and the user may have been unaware that the alarms were off.

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

Philips

Manufacturer

Not provided

Phone

Not provided

Brand: Philips.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

critical

Recall details

Recall number

81234

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Hard disk drives in azurion and allura systems

Critical riskHealth Canada15 May 2026
CA
Recall image placeholder

Ingenia and achieva systems

Critical riskHealth Canada16 Apr 2026
CA
Recall image placeholder

Aneurysmflow

Medium riskHealth Canada2 Apr 2026
CA
Recall image placeholder

Intellivue mx patient monitor system-intellibridge module ec10

Medium riskHealth Canada11 Mar 2026
CA
Recall image placeholder

Azurion systems

Critical riskHealth Canada7 May 2026
CA
Recall image placeholder

Azurion and allura xper systems

Medium riskHealth Canada6 May 2026
CA
Recall image placeholder

Allura xper and azurion systems

Medium riskHealth Canada17 Mar 2026
CA
Recall image placeholder

Philips allura xper systems

Medium riskHealth Canada4 Jun 2026
CA
Recall image placeholder

Trilogy evo and ev300

High riskHealth Canada18 Feb 2026
CA
Recall image placeholder

Allura xper and allura cv20 systems

Critical riskHealth Canada6 Jan 2026
CA
Recall image placeholder

Azurion r2.1.10 and r2.2.10 systems

Critical riskHealth Canada29 Dec 2025
CA
Recall image placeholder

Zenition 50

Medium riskHealth Canada15 Oct 2025

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 20 Feb 2026. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.