Open official notice

Oxford Partial Knee System, Fixed Lateral Tibial Construct- Intended to replace part of a knee joint when used in conjunction with a

Official notice

1 Feb 2023

RecallScope risk estimate:MediumFDA

An official FDA notice concerns Oxford Partial Knee System, Fixed Lateral Tibial Construct- Intended to replace part of a knee joint when used in conjunction with a. Packaged in the incorrect shelf carton. Incorrect shelf carton; outer carton is smaller than the required outer carton, which forces the Tyvex pouch to be folded and packaged in the carton causing compound creases and compromise the sterile barrier Published 2023-02-01.

What to do now

Stop using the affected device and follow the corrective action in the official FDA notice. Contact your healthcare provider or the manufacturer for next steps.

Units affected

7

Recall number

Z-1018-2023

Official notice

2023-02-01

The hazard

Packaged in the incorrect shelf carton. Incorrect shelf carton; outer carton is smaller than the required outer carton, which forces the Tyvex pouch to be folded and packaged in the carton causing compound creases and compromise the sterile barrier

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

077830

Event Ids

91365

Recall Numbers

Z-1018-2023

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

Oxford Partial Knee System, Fixed Lateral Tibial Construct- Intended to replace part of a knee joint when used in conjunction with a femoral and tibial component Item Number: 154341

Units affected

7

Hazard

Packaged in the incorrect shelf carton. Incorrect shelf carton; outer carton is smaller than the required outer carton, which forces the Tyvex pouch to be folded and packaged in the carton causing compound creases and compromise the sterile barrier

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Affected area 2

CA · Canada

Affected area 3

JP · Japan

Affected area 4

NL · Netherlands

Affected area 5

NZ · New Zealand

Company & contacts

Brand

Manufacturer

Biomet, Inc.

Phone

Not provided

Brand: Biomet. Contact: Biomet, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-1018-2023

Event ID

91365

Units affected

7

Status

active

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 1 Feb 2023. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button