(1) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Bi-Polar Cup, E1 Antioxidant Infused, 28 MM, 41 MM: Item No. 110010458

Official notice

21 Oct 2020

RecallScope risk estimate:MediumFDA

An official FDA notice concerns (1) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Bi-Polar Cup, E1 Antioxidant Infused, 28 MM, 41 MM: Item No. 110010458. Zimmer Biomet is conducting a medical device recall for three lots of the RingLoc Bi-Polar Hip system Acetabular Cup. The affected lots were gamma sterilized twice. Sufficient data does not exist to support the functionality, shelf life, or package integrity for more than one-time sterilization. Published 2020-10-21.

What to do now

This recall has been terminated. See the official FDA notice for the latest status.

Units affected

48

Recall number

Z-0097-2021

Official notice

2020-10-21

The hazard

Zimmer Biomet is conducting a medical device recall for three lots of the RingLoc Bi-Polar Hip system Acetabular Cup. The affected lots were gamma sterilized twice. Sufficient data does not exist to support the functionality, shelf life, or package integrity for more than one-time sterilization.

How to check if you’re affected

Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.

Lots

710930, 649750, 433290

Event Ids

86382

Recall Numbers

Z-0097-2021

Checked manually this time?

Follow brands and keywords in the app to receive alerts for matching official notices.

Turn on recall alerts

Affected products

Product 1

(1) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Bi-Polar Cup, E1 Antioxidant Infused, 28 MM, 41 MM: Item No. 110010458; Lot No. 710930; UDI No. (01) 00880304568747 (17) 230614 (10) 710930 (2) RingLoc Hip system Acetabular Bi-Polar Cup: Acetabular Cup, ArCom, 41 MM OD, 28 MM ID: Item No. 11-165206, Lot No. 649750; UDI No. (01) 00880304001923 (17) 230614 (10) 649750 (3) RingLoc Bi-Polar Hip System Acetabular Cup ArCom: Acetabular Cup, ArCom, 52 MM OD, 28 MM ID: Item No. 11-165228; Lot No. 433290; UDI No. (01) 00880304001930 (17) 230612 (10) 433290

Units affected

48

Hazard

Zimmer Biomet is conducting a medical device recall for three lots of the RingLoc Bi-Polar Hip system Acetabular Cup. The affected lots were gamma sterilized twice. Sufficient data does not exist to support the functionality, shelf life, or package integrity for more than one-time sterilization.

Where it was sold

Retailer information is not currently available in RecallScope. Check the official notice for distribution details.

Location 1

US · United States

Affected area 1

US · United States

Company & contacts

Brand

Manufacturer

Biomet, Inc.

Phone

Not provided

Brand: Biomet. Contact: Biomet, Inc..

Source & status

Source

Tier

official

Category

Tech & Electronics

Official classification

Class II

Recall details

Recall number

Z-0097-2021

Event ID

86382

Units affected

48

Status

terminated

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Weekly recall digest

Get next week’s recalls before you need them.

One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest arrives once a week. For matching recall alerts after notices are processed, get the app.

This recall was published 21 Oct 2020. The next one won’t wait for you to search.

RecallScope checks listed publishers for the brands and VINs you follow, and can alert you when a matching notice is processed.

apple-buttongoogle-button