Image not available in RecallScope.
Arcos Modular Revision Hip System, Standard Cone Prox Body, Porous Plasma,Size A, 60 MM Item Number: 11-301301
Official notice
2 Mar 2022
An official FDA notice concerns Arcos Modular Revision Hip System, Standard Cone Prox Body, Porous Plasma,Size A, 60 MM Item Number: 11-301301. The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor. Published 2022-03-02.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
19
Recall number
Z-0670-2022
Official notice
2022-03-02
The hazard
The hazard
The products in scope underwent an incorrect rework operation for the porous plasma sprayed coating. Internal testing indicated that the parts may not have sufficient adhesion strength. The scope was identified to be all lots that were porous plasma coated or recoated at a vendor that were then reworked through the internal waterjet rework process. The waterjet rework process for porous plasma coating is not validated for products that were coated at a vendor.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Lots
Event Ids
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Affected area 2
Affected area 3
Company & contacts
Company & contacts
Brand: Biomet. Contact: Biomet, Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest