Image not available in RecallScope.
Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Reconstruction, Distal Humeral
Official notice
4 Apr 2018
An official FDA notice concerns Zimmer Biomet Comprehensive Segmental Revision System (SRS). Orthopedic implants for Proximal Humeral Reconstruction, Distal Humeral. During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Condyle affects assembly during surgery. If the product has already been implanted, no action is required. The mismatch is estimated to prevent screw engagement in less than 1% of cases. Published 2018-04-04.
What to do now
This recall has been terminated. See the official FDA notice for the latest status.
Units affected
634
Recall number
Z-1173-2018
Official notice
2018-04-04
The hazard
The hazard
During assembly, an SRS Humeral Distal Body implant failed to engage correctly at the point of reduction with a Discovery Elbow System Humeral Condyle. A mismatch between the screw holes in the Humeral Body & Elbow Condyle affects assembly during surgery. If the product has already been implanted, no action is required. The mismatch is estimated to prevent screw engagement in less than 1% of cases.
How to check if you’re affected
How to check if you’re affected
Compare the product name and identifiers below with the item you own. Check the official notice for the affected products and next steps.
Lots
Skus
Event Ids
Model numbers
Recall Numbers
Checked manually this time?
Follow brands and keywords in the app to receive alerts for matching official notices.
Turn on recall alertsAffected products
Affected products
Product 1
Units affected
Hazard
Where it was sold
Where it was sold
Retailer information is not currently available in RecallScope. Check the official notice for distribution details.
Location 1
Affected area 1
Company & contacts
Company & contacts
Brand: Biomet. Contact: Biomet, Inc..
Source & status
Source & status
Recall details
Recall details
Recall number
Event ID
Units affected
Status
Source tier
Documents, notices & source
See the official notice and source documents for full details.
View full official noticeWeekly recall digest
Get next week’s recalls before you need them.
One email every Monday with official-source recalls for your country, and nothing else. Unsubscribe from any email.
The digest arrives once a week. For matching recall alerts after notices are processed, get the app.