Merlin patient care system - programmer software

Official notice

27 Jun 2022

Last checked

27 Jun 2022

Medium riskHealth Canada
Watch St Jude Medical Cardiac Rhythm Management Division

An official Health Canada notice concerns Merlin patient care system - programmer software. Abbott is notifying customers of the potential for programmer software (Merlin pcs) and remote monitoring software applications (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers. Pacemaker/battery functionality, therapy delivery, and longevity remain normal and within specifications. Voltage measurements and ERI (elective replacement indicator), which is based on direct voltage measurement, remain accurate. Recall start date: June 16, 2022 Published 2022-06-27.

What to do now

Contact the manufacturer if you require additional information.

Recall number

64306

Official notice

2022-06-27

The hazard

Abbott is notifying customers of the potential for programmer software (Merlin pcs) and remote monitoring software applications (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers. Pacemaker/battery functionality, therapy delivery, and longevity remain normal and within specifications. Voltage measurements and ERI (elective replacement indicator), which is based on direct voltage measurement, remain accurate. Recall start date: June 16, 2022

How to check if you're affected

Compare the VIN on your vehicle against the official list. If your VIN is listed, follow the official notice.

Recall Numbers

64306

Checked manually this time?

The app watches your VIN and alerts you automatically next time — no searching, no checking back.

Turn on automatic VIN checks

Affected products

Product 1

Affected products - Merlin Patient Care System - Programmer Software, Lot or serial number - version v25.3.x or earlier, Model or catalogue number - 3330 VER 25.2.4 3330 VER 24.4.4 3330 VER 18.2 3330 VER 24.6.4 3330 VER 24.3.4 3330 VER 17.2.1 3330 VER 23.0.4 3330 VER 25.0.4 3330 VER 25.3.4 3330 VER 24.2.4 3330 VER 20.0.1.2 3330 VER 24.5.4

Risk / hazard

Abbott is notifying customers of the potential for programmer software (Merlin pcs) and remote monitoring software applications (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers. Pacemaker/battery functionality, therapy delivery, and longevity remain normal and within specifications. Voltage measurements and ERI (elective replacement indicator), which is based on direct voltage measurement, remain accurate. Recall start date: June 16, 2022

Where it was sold

The source does not list specific retailers. See the official notice for where affected vehicles were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Brand

St Jude Medical Cardiac Rhythm Management Division, Merlin Patient Care System - Programmer Software

Manufacturer

Not provided

Phone

Not provided

Brand: St Jude Medical Cardiac Rhythm Management Division, Merlin Patient Care System - Programmer Software.

Source & status

Source

Health Canada

Tier

official

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

64306

Classification

Type II

Source tier

official

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

CA
Recall image placeholder

Merlin™ 3650 patient care system (pcs)

Medium riskHealth Canada22 May 2026
CA
Recall image placeholder

Assurity and endurity pacemakers

Medium riskHealth Canada27 Feb 2025
CA
Recall image placeholder

Merlin pcs 3330 software upgrade kit

Medium riskHealth Canada6 Dec 2024
CA
Recall image placeholder

Neutrino nxt implantable cardioverter defibrilator (icds) and cardiac resynchronization therapy defibrillator (crt-ds)

Medium riskHealth Canada15 Sept 2023
CA
Recall image placeholder

Assurity mri pulse generator

High riskHealth Canada2 Aug 2022
CA
Recall image placeholder

Merlin patient care system - programmer software

Medium riskHealth Canada29 Mar 2022
CA
Recall image placeholder

Heartsine® samaritan® pad sam 350p sam 360p sam 500p - membrane tail

Medium riskHealth Canada23 Jul 2026
CA
Recall image placeholder

Abiomed introducer kits with impella heart pumps

High riskHealth Canada29 Jun 2026
CA
Recall image placeholder

Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Medium riskHealth Canada22 Jun 2026
CA
Recall image placeholder

Faradrive™ steerable sheath

Medium riskHealth Canada19 Jun 2026
CA
Recall image placeholder

Artis icono ceiling

Medium riskHealth Canada17 Jun 2026
CA
Recall image placeholder

Ge healthcare integrated and stand-alone resuscitation systems

Medium riskHealth Canada16 Jun 2026

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 27 Jun 2022. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.