Merlin™ 3650 patient care system (pcs)

Official notice

22 May 2026

MediumHealth Canada

An official Health Canada notice concerns Merlin™ 3650 patient care system (pcs). Abbott is reaching out to customers to inform about a Merlin™ PCS 3650 programmer software update that addresses a software behavior that may occur during Pacing Capture Threshold (PCT) test of an AVEIR™ Leadless Pacemaker (LP). If intermittent loss of telemetry occurs during PCT test, the programmer may not successfully communicate the command to terminate the test, potentially leaving the LP in sub-threshold pacing until telemetry is fully disconnected. This issue can only occur during a PCT test, which is performed during implant or in clinic follow-up. This issue was identified through a review of complaints data, which confirmed four (4) complaints associated with the issue globally. Published 2026-05-22.

What to do now

Contact the manufacturer if you require additional information.

Recall number

82105

Official notice

2026-05-22

The hazard

Abbott is reaching out to customers to inform about a Merlin™ PCS 3650 programmer software update that addresses a software behavior that may occur during Pacing Capture Threshold (PCT) test of an AVEIR™ Leadless Pacemaker (LP). If intermittent loss of telemetry occurs during PCT test, the programmer may not successfully communicate the command to terminate the test, potentially leaving the LP in sub-threshold pacing until telemetry is fully disconnected. This issue can only occur during a PCT test, which is performed during implant or in clinic follow-up. This issue was identified through a review of complaints data, which confirmed four (4) complaints associated with the issue globally.

How to check if you're affected

Compare the product name and identifiers below with the item you own. Follow the official notice if they match.

Skus

3650

Other

RA RA-82105

Model Numbers

3650

Recall Numbers

82105

Checked manually this time?

The app alerts you when a new official recall is published — no searching, no checking back.

Turn on recall alerts

Affected products

Product 1

Lot or serial number - Not applicable., Model or catalogue number - 3650

Product 2

Lot or serial number - Not applicable., Model or catalogue number - 3330 V28.1.4

Product 3

Lot or serial number - Not applicable., Model or catalogue number - 3330 VER 25.2.4

Product 4

Lot or serial number - Not applicable., Model or catalogue number - 3330 V25.8.4

Product 5

Lot or serial number - Not applicable., Model or catalogue number - 3330 V26.0.4

Product 6

Lot or serial number - Not applicable., Model or catalogue number - 3330 V27.0.4

Product 7

Lot or serial number - Not applicable., Model or catalogue number - 3330 V28.8.4

Product 8

Lot or serial number - Not applicable., Model or catalogue number - 3330 V25.4.4

Product 9

Lot or serial number - Not applicable., Model or catalogue number - 3330 V28.3.4

Product 10

Lot or serial number - Not applicable., Model or catalogue number - 3330 V28.2.4

Product 11

Lot or serial number - Not applicable., Model or catalogue number - 3330 VER 25.0.4

Product 12

Lot or serial number - Not applicable., Model or catalogue number - 3330 VER 25.3.4

Product 13

Lot or serial number - Not applicable., Model or catalogue number - 3330 V28.0.4

Product 14

Lot or serial number - Not applicable., Model or catalogue number - 3330 V28.5.4

Risk / hazard

Abbott is reaching out to customers to inform about a Merlin™ PCS 3650 programmer software update that addresses a software behavior that may occur during Pacing Capture Threshold (PCT) test of an AVEIR™ Leadless Pacemaker (LP). If intermittent loss of telemetry occurs during PCT test, the programmer may not successfully communicate the command to terminate the test, potentially leaving the LP in sub-threshold pacing until telemetry is fully disconnected. This issue can only occur during a PCT test, which is performed during implant or in clinic follow-up. This issue was identified through a review of complaints data, which confirmed four (4) complaints associated with the issue globally.

Where it was sold

The source does not list specific retailers. See the official notice for where affected products were sold.

Location 1

CA · Canada

Affected area 1

CA · Canada

Company & contacts

Manufacturer

Not provided

Phone

Not provided

Brand: St Jude Medical Cardiac Rhythm Management Division.

Source & status

Source

Health Canada

Tier

beta

Category

HEALTH

Classification

Type II

App estimate

medium

Recall details

Recall number

82105

Classification

Type II

Source tier

beta

Documents, notices & source

See the official notice and source documents for full details.

View full official notice
Something wrong on this page?

Related recalls

HEALTH
Medium

Merlin pcs 3330 software upgrade kit

Health Canada6 Dec 2024

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Neutrino nxt implantable cardioverter defibrilator (icds) and cardiac resynchronization therapy defibrillator (crt-ds)

Health Canada15 Sept 2023

Brand match

Medium risk

Markets

CA
HEALTH
High

Assurity mri pulse generator

Health Canada2 Aug 2022

Brand match

High risk

Markets

CA
HEALTH
Medium

Merlin patient care system - programmer software

Health Canada27 Jun 2022

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Merlin patient care system - programmer software

Health Canada29 Mar 2022

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Assurity and endurity pacemakers

Health Canada27 Feb 2025

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Heartsine® samaritan® pad sam 350p sam 360p sam 500p - membrane tail

Health Canada23 Jul 2026

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Intramedullary bone saw blade assembly

Health Canada3 Jul 2026

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Glidesheath slender® introducer sheath

Health Canada2 Jul 2026

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Hoffmann® ii carbon connecting rod

Health Canada23 Jun 2026

Brand match

Medium risk

Markets

CA
HEALTH
Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)Medium

Importation of chinese-authorized ifosfamide for injection due to the current shortage of canadian-authorized ifex (ifosfamide for injection)

Health Canada22 Jun 2026

Brand match

Medium risk

Markets

CA
HEALTH
Medium

Faradrive™ steerable sheath

Health Canada19 Jun 2026

Brand match

Medium risk

Markets

CA

Weekly recall digest

Get next week's recalls before you need them.

One email every Monday — official-source recalls for your country, and nothing else. Unsubscribe from any email.

The digest is the slow lane — one email a week. For instant, severity-gated alerts the moment a notice is published, get the app.

This recall was published 22 May 2026. The next one won't wait for you to search.

RecallScope watches official sources for the brands and VINs you follow, and alerts you the moment a new notice is published.